Figure 2-8: Relationship Between TIC and pH

Total Organic Carbon Analyzer

Theory of Operation

 

 

 

be reduced to approximately 2.0 to 3.0. If the sample pH is not properly reduced as described in Figure 2-8, increase the normalcy of the acid reagent until the water exiting the sparger drain is between 2.0 and 3.0.

Figure 2-8: Relationship Between TIC and pH

2.4.3 Oxidation

The UV reactor and persulfate reagent provide oxidation of the carbon to CO2. The UV reactor consists of the following:

Mounting enclosure with viewing window

Borosilicate reactor body

Heater/Temperature sensor

UV Lamp

Miscellaneous fittings and clamps