Sample requirements:

Appendix

Model 6750

 

 

Sample requirements:

Inlet Pressure: Atmospheric to 3 psig

Flow Rate: 20cc/min. (nom)

Drain: Gravity drain vented to atmosphere

Suspended Solids: 1,000 microns (max.)

Reagents (TOC Mode): Sodium Persulfate: 4.0 gal/month Phosphoric Acid, 10% v/v: 4.0 gal/month (nom.).

Calibration: One point span, Chemical standard, computer-stored multiple calibration curves.

Facility requirements:

Power: 115 + 10% VAC, 50/60 Hz 7A Service Recommended.

220 + 10% VAC, 50/60 Hz (optional) 4 A Service Recommended

Air: Clean, dry, oil free, CO2 free instrument air at 15 psi +/- 2 psi, consumption < 500 cc/minute.