Total Organic Carbon Analyzer | Appendix |
|
|
Appendix
A.1Specifications
Measurement Methods: TC (Total Carbon) – UV/Persulfate
Oxidation
TOC (Total Organic Carbon) acidification and sparging to eliminate inorganic carbon interference.
Temperature: 25o C ± 5o
Measurement Ranges:
Note: Range changes often require carrier gas adjustments. The Model 6750 uses the precision of computer controlled mass flow controllers to eliminate operator error and the use of inaccurate mechanical flow meters.
Display: | Windows with Paperless Chart Recorder |
| 2 Line LCD |
Data Handling: | |
Analog Output: | |
Alarms: | Two concentration alarms |
| One master fault alarm |
Performance:
Response Time: Dependant upon application
Repeatability: +2% FS (std. Dev. 1 sigma)
Zero/span stability: +2% FS (at 25o C)
Linearity: +2% FS
Teledyne Analytical Instruments | 69 |