Total Organic Carbon Analyzer

Appendix

 

 

Appendix

A.1Specifications

Measurement Methods: TC (Total Carbon) – UV/Persulfate

Oxidation

TOC (Total Organic Carbon) acidification and sparging to eliminate inorganic carbon interference.

Temperature: 25o C ± 5o

Measurement Ranges: 0-10 through 0-10,000 ppm, full scale (std.)

Note: Range changes often require carrier gas adjustments. The Model 6750 uses the precision of computer controlled mass flow controllers to eliminate operator error and the use of inaccurate mechanical flow meters.

Display:

Windows with Paperless Chart Recorder

 

2 Line LCD

Data Handling:

RS-232C, RS-485

Analog Output:

4-20mA

Alarms:

Two concentration alarms

 

One master fault alarm

Performance:

Response Time: Dependant upon application

Repeatability: +2% FS (std. Dev. 1 sigma)

Zero/span stability: +2% FS (at 25o C)

Linearity: +2% FS

Teledyne Analytical Instruments

69

Page 81
Image 81
Teledyne 6750 operating instructions 1Specifications, Total Organic Carbon Analyzer Appendix