Teledyne 6750 operating instructions b.Select TOC The following screen appears

Models: 6750

1 96
Download 96 pages 40.67 Kb
Page 51
Image 51
b.Select [TOC].

Total Organic Carbon Analyzer

Setup and Operation

 

 

 

Note: On the initial Startup, IR gas calibration should be performed prior to liquid calibration

a.Select [CALIBRATE] from the main menu. The following screen will appear:

b.Select [TOC].

The following screen appears:

c.Flow “ZERO” (D. I. Water) and allow NDIR reading to stabilize. For maximum accuracy, especially at lower ranges, this process may take 20 minutes. When reading is stable, select [ZERO].

d.Flow Liquid Span Standard Solution. Allow NDIR reading to stabilize, (approximately 15 minutes).

e.Select [SPAN]. It is recommended that the operator continue to observe the reading and if not stabilized at the desired span setting ± 2%, repeat the span selection until stable.

Teledyne Analytical Instruments

39

Page 51
Image 51
Teledyne 6750 operating instructions b.Select TOC The following screen appears