Total Organic Carbon Analyzer | Maintenance |
|
|
11.Position the other end bell on top and press the end bells together.
12.Be careful not to pinch the tubing.
13.If end bells do not snap tightly together, reload tubing.
14.If necessary, turn key in slot on rotor shaft to adjust tubing (as in next step - E).
15.With key in slot on rotor shaft, turn key to align tang on rotor shaft with slot in motor drive shaft.
16.Point tubing retainer grooves up.
17.Shift the pump head slightly until it snaps on the alignment pins (if present).
18.Secure with four provided screws.
19.Tighten with fingers only.
Figure 5-15: Pump Head Tubing Replacement
5.3.134-20 mA Adjustment Procedure
1.Select [TEST] from the Run Screen task bar, then, select
[SYSTEM DIAGNOSTICS].
2.Locate Analog A cursor and position the cursor to MAX for a moment, then, return the cursor to MIN.
Teledyne Analytical Instruments | 67 |