Maintenance

Model 6750

 

 

From this screen the operator may perform individual testing of components related to functionality and settings.

5.2 Troubleshooting

If you are experiencing trouble with your instrument, refer to Table 5-1: Troubleshooting.

Table 5-1: Troubleshooting

PROBLEMS

Analyzer will not turn on

SOLUTION

Check fuse & power connections.

Look for loose or damaged wiring.

Check for 12V DC from power supply.

Check power supply fuse.

Analyzer does not Calibrate, is

Look for & correct any liquid or gas line

sluggish

leaks.

 

Analyzer is excessively noisy

Display not functioning

Check for leaks at tubing connections

Check cable connections

Check power supply voltage

Suspect:

Malfunctioning Computer (P/N ST13039).

Master Control Panel Assembly (P/N ST13042-1)

Pump head not rotating

Pump head rotating but not pumping fluid

Refer to section “Pumps” 12.3.12

Verify that carrier loss did not cause the pumps to be turned off.

Check for AC voltage at terminal strip on inside of application panel.

Refer to section “PUMPS”

Change tubing

Verify sample line is in liquid.

Teledyne Analytical Instruments

44

Page 56
Image 56