Maintenance | Model 6750 |
|
|
From this screen the operator may perform individual testing of components related to functionality and settings.
5.2 Troubleshooting
If you are experiencing trouble with your instrument, refer to Table
Table 5-1: Troubleshooting
PROBLEMS
Analyzer will not turn on
SOLUTION
•Check fuse & power connections.
•Look for loose or damaged wiring.
•Check for 12V DC from power supply.
•Check power supply fuse.
Analyzer does not Calibrate, is | • Look for & correct any liquid or gas line |
sluggish | leaks. |
|
Analyzer is excessively noisy
Display not functioning
•Check for leaks at tubing connections
•Check cable connections
•Check power supply voltage
•Suspect:
•Malfunctioning Computer (P/N ST13039).
•Master Control Panel Assembly (P/N
Pump head not rotating
Pump head rotating but not pumping fluid
•Refer to section “Pumps” 12.3.12
•Verify that carrier loss did not cause the pumps to be turned off.
•Check for AC voltage at terminal strip on inside of application panel.
•
•Refer to section “PUMPS”
•Change tubing
•Verify sample line is in liquid.
Teledyne Analytical Instruments | 44 |