Teledyne Unpacking the BioOptix 10 System, BioOptix 10 Installation, Space, Requirements

Page 19

BioOptix™ 10

Section 2 Installation

2.1Unpacking the BioOptix 10 System

This section provides instructions for installing the system and preparing it for operation.

The BioOptix 10 system consists of a BioOptix 10 unit, a Foxy 200 Fraction Collector, and a Windows PC running BioOptix 10 software.

WARNING

The system is heavy. Use a two-person lift to prevent injury.

Carefully unpack the shipment and inspect the contents to verify receipt of all components against the packing list. If there is any damage to the shipping carton or any components, contact the shipping agent and Teledyne Isco (or its authorized represen- tative) immediately.

WARNING

If there is any evidence that the system has been damaged in shipping, do not plug it into the power line. Contact Teledyne Isco or its authorized representative for advice.

Compare the contents of the boxes with the enclosed packing slip. If there are any shortages, contact Teledyne Isco immediately.

2.2BioOptix 10 Installation

2.2.1BioOptix 10 Location The system should be installed on a work surface capable of sup-

and Space

porting the system, and buffer, sample, and waste containers.

Requirements

The system is designed for ambient temperature and cold-room

 

operation. Refer to Table 1-1 for environmental specifications.

 

The BioOptix 10 unit requires about 3350 square centimeters

 

(520 in2) of bench space. Space requirements are shown in Figure

 

2-1. Space must be left behind the unit for case ventilation and

 

electrical connections. The left side of the unit should be kept

 

clear to access the power switch and for fluid connections.

CAUTION

Ensure there is at least 10 cm (4 inches) of air space on its left side so that the disconnect device (power switch) may be eas- ily accessed in an emergency.

2-1

Image 19
Contents BioOptix Page Foreword Page Application, take heed in the following general warnings General WarningsDamage Symboles de sécurité Hazard SymbolsBioOptix 10 Safety BioOptix Table of ContentsAppendix a Method Templates System MaintenanceAppendix B Fluid Path Diagrams Appendix C Replacement Parts ListsList of Figures List of Tables Product Overview IntroductionMaximum Altitude 2000 meters BioOptix 10 SpecificationsBioOptix 10 Front Panel Controls System Controls and ConnectorsBioOptix 10 Front Panel Controls BioOptix 10 Left-side Controls and Connections BioOptix 10 Left-side ConnectionsBioOptix 10 Right-side Connections BioOptix 10 Right-side ConnectionsBioOptix 10 Rear Panel Controls and Connections BioOptix 10 Rear Panel Controls and ConnectionsSpace InstallationBioOptix 10 Installation Unpacking the BioOptix 10 SystemAssembling ElectricalBioOptix Accessing pump and column module internal connections Routing the Sample Out lines column module P110 Connection on BD3 AC power connection from pump module 10 Liquid connections pump module shown 11 Aligning the fasteners Foxy Ceeding with the arm assembly instructionsBioOptix Installation Common Buffer Inlet Mounting Block Manifold Channel Numbers Color Part Number Tubing Color chart16 Foxy 200 tubing connections Windows PC Minimum Requirements ConfiguringBioOptix 10 Software BioOptix 10 System Preparation 18 Column bypass tubing 20 BioOptix 10 Manual Hardware Control window BioOptix Installation Operation Sample Preparation Column InstallationAdapters Gradients BioOptix 10 SoftwareOverview Defining a Gradient To define a gradient MethodsEquilibrate tab settings Load tab settings Load Settings This block loads the sample onto the columnWash Settings Experiments Flush SettingsSetting up File RunningExperiment Operation ButtonsOperation Zoom and panPump Flush Post-experiment OperationPreview Saving Run DataAutoReports BioOptix 10 Operation Vent damage System MaintenanceExterior Cleaning Piston Seal Replacement Refer to -3 for the following steps Pump assembly components Deuterium Lamp Replacement Waste port mounting plate Common Problems TroubleshootingOperating Problems Inaccessible Fuses Circuit breaker location How to Ship Returns Service DepartmentTeledyne Isco, Inc Appendix a Method Templates Dard.tpl General Editing Rules Modifying TemplateBlockn BlocksOther Fluid-type Blocks Appendix B Fluid Path DiagramsLoad-type Blocks BioOptix Appendix B Fluid Path DiagramsFigure B-4 Fluid path for Load-type blocks Manual injections Pump Flush Plumbing DiagramFigure B-7 BioOptix 10 Plumbing Diagram BioOptix Appendix B Fluid Path Diagrams Appendix C Replacement Parts Lists BioOptix Appendix C Replacement Parts Lists Detector Assembly 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Column Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Pump Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Declaration of Conformity Page Teledyne Isco One Year Limited Factory Service Warranty