Teledyne 10 manual Foxy, Ceeding with the arm assembly instructions

Page 27

BioOptix™ 10

Section 2 Installation

12.Ensure the Power Switch is in the Off (“0”) position. Then, connect the AC power cord to the back of the Pump Mod- ule. Connect the other end of the power cord to an outlet meeting the electrical requirements listed on the serial number label.

13.Locate the Serial “Y” cable, P/N 60-1020-241 in the Acces- sories Kit. Connect the 25-pin connector labeled First Instrument Optix to the External Control port on the back of the Pump Module.

The BioOptix 10 may now be positioned on the bench as shown in Figure 2-1.

WARNING

The system is heavy. Use a two-person lift to prevent injury.

 

14.

Insert the ten column clamps into the BioOptix 10. Secure

 

 

the clamps by tighten the thumbscrews against the stain-

 

 

less steel shafts (Figure 2-12).

2.2.4 Assembling the

1.

Refer to section 2 of the Foxy 200 X-Y Fraction Collector

Foxy 200

 

instruction manual. Unpack the Foxy 200 and assemble

 

 

the arm.

 

 

 

 

 

Note

 

To prevent damage to the ground wire on the fraction collector

 

arm, attach the ten-channel drop former to the arm before pro-

 

ceeding with the arm assembly instructions.

 

2.

Loosen the two thumbscrews and raise the ten-channel

 

 

drop former assembly to its highest position (Figure 2-13).

 

 

Retighten the thumbscrews.

Drop former height thumbscrews

10-channel drop former assembly

Figure 2-13 Foxy arm

3.Load collection tubes into the racks. Place the racks on the Foxy 200.

4.Loosen the two thumbscrews and lower the ten-channel drop former assembly so that it will clear the collection tubes (Figure 2-13). Retighten the thumbscrews.

2-9

Image 27
Contents BioOptix Page Foreword Page Application, take heed in the following general warnings General WarningsDamage Symboles de sécurité Hazard SymbolsBioOptix 10 Safety BioOptix Table of ContentsAppendix a Method Templates System MaintenanceAppendix B Fluid Path Diagrams Appendix C Replacement Parts ListsList of Figures List of Tables Product Overview IntroductionMaximum Altitude 2000 meters BioOptix 10 SpecificationsSystem Controls and Connectors BioOptix 10 Front Panel ControlsBioOptix 10 Front Panel Controls BioOptix 10 Left-side Controls and Connections BioOptix 10 Left-side ConnectionsBioOptix 10 Right-side Connections BioOptix 10 Right-side ConnectionsBioOptix 10 Rear Panel Controls and Connections BioOptix 10 Rear Panel Controls and ConnectionsSpace InstallationBioOptix 10 Installation Unpacking the BioOptix 10 SystemElectrical AssemblingBioOptix Accessing pump and column module internal connections Routing the Sample Out lines column module P110 Connection on BD3 AC power connection from pump module 10 Liquid connections pump module shown 11 Aligning the fasteners Foxy Ceeding with the arm assembly instructionsBioOptix Installation Common Buffer Inlet Mounting Block Manifold Channel Numbers Color Part Number Tubing Color chart16 Foxy 200 tubing connections Configuring Windows PC Minimum RequirementsBioOptix 10 Software BioOptix 10 System Preparation 18 Column bypass tubing 20 BioOptix 10 Manual Hardware Control window BioOptix Installation Sample Preparation Column Installation OperationAdapters BioOptix 10 Software GradientsOverview Defining a Gradient To define a gradient MethodsEquilibrate tab settings Load tab settings Load Settings This block loads the sample onto the columnWash Settings Experiments Flush SettingsSetting up File RunningExperiment Operation ButtonsOperation Zoom and panPump Flush Post-experiment OperationPreview Saving Run DataAutoReports BioOptix 10 Operation Vent damage System MaintenanceExterior Cleaning Piston Seal Replacement Refer to -3 for the following steps Pump assembly components Deuterium Lamp Replacement Waste port mounting plate Troubleshooting Common ProblemsOperating Problems Inaccessible Fuses Circuit breaker location Service Department How to Ship ReturnsTeledyne Isco, Inc Appendix a Method Templates Dard.tpl General Editing Rules Modifying TemplateBlocks BlocknOther Fluid-type Blocks Appendix B Fluid Path DiagramsLoad-type Blocks BioOptix Appendix B Fluid Path DiagramsFigure B-4 Fluid path for Load-type blocks Manual injections Pump Flush Plumbing DiagramFigure B-7 BioOptix 10 Plumbing Diagram BioOptix Appendix B Fluid Path Diagrams Appendix C Replacement Parts Lists BioOptix Appendix C Replacement Parts Lists Detector Assembly 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Column Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Pump Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Declaration of Conformity Page Teledyne Isco One Year Limited Factory Service Warranty