BioOptix™ 10

 

Section 3

Operation

 

 

 

 

3.6.1

Setting up the

Before proceeding, determine which BioOptix 10 channel posi-

 

Experiment

tions will be used during the purification run. Channels that will

 

 

be in use require that the corresponding rows in the table be

completed. Note that to use these positions they should have samples prepared (section 3.1) and columns installed (3.2).

Table rows for unused channels must remain empty.

To set up the experiment:

1. Position the cursor in the Sample Name table cell of the first channel position in use. Enter a descriptive name for the sample. This activates the channel for use in the exper- iment.

Because the Sample Name also is used as a filename for post-run data, the text you enter in this field must be a Windows-compatible file name.

2. Advance the cursor to the next cell to the right.

BioOptix 10 completes the remaining cells in the row using the settings from the method file as defaults.

3. Repeat steps 1 and 2 for all channel positions that will be in use.

The table now is populated with default settings for each channel position in use.

Figure 3-7 Experiment File

3.6.2Editing the Experiment The table can be edited to create unique purification parameters for each channel. To edit any cell in the table, position the cursor in a cell and enter the new value. You can also select different gradient files by clicking on the [...] button and browsing for the new file. When editing the table note the following:

Load type values (such as the “Load Column in Figure 3-7) must be a multiple of the stroke volume. Stroke volume is discussed in section 2.3.1.

If selecting a different gradient file, note that the total volume pumped be the same for all channels in the experiment.

3-8

Image 44
Contents BioOptix Page Foreword Page General Warnings Application, take heed in the following general warningsDamage Hazard Symbols Symboles de sécuritéBioOptix 10 Safety Table of Contents BioOptixSystem Maintenance Appendix B Fluid Path DiagramsAppendix C Replacement Parts Lists Appendix a Method TemplatesList of Figures List of Tables Introduction Product OverviewBioOptix 10 Specifications Maximum Altitude 2000 metersBioOptix 10 Front Panel Controls System Controls and ConnectorsBioOptix 10 Front Panel Controls BioOptix 10 Left-side Connections BioOptix 10 Left-side Controls and ConnectionsBioOptix 10 Right-side Connections BioOptix 10 Right-side ConnectionsBioOptix 10 Rear Panel Controls and Connections BioOptix 10 Rear Panel Controls and ConnectionsInstallation BioOptix 10 InstallationUnpacking the BioOptix 10 System SpaceBioOptix ElectricalAssembling Accessing pump and column module internal connections Routing the Sample Out lines column module P110 Connection on BD3 AC power connection from pump module 10 Liquid connections pump module shown 11 Aligning the fasteners Ceeding with the arm assembly instructions FoxyBioOptix Installation Common Buffer Inlet Mounting Block Manifold Tubing Color chart Channel Numbers Color Part Number16 Foxy 200 tubing connections BioOptix 10 Software ConfiguringWindows PC Minimum Requirements BioOptix 10 System Preparation 18 Column bypass tubing 20 BioOptix 10 Manual Hardware Control window BioOptix Installation Adapters Sample Preparation Column InstallationOperation Overview BioOptix 10 SoftwareGradients Methods Defining a Gradient To define a gradientEquilibrate tab settings Load Settings This block loads the sample onto the column Load tab settingsWash Settings Flush Settings ExperimentsSetting up Running ExperimentOperation Buttons FileZoom and pan OperationPost-experiment Operation Pump FlushSaving Run Data PreviewAutoReports BioOptix 10 Operation System Maintenance Vent damageExterior Cleaning Piston Seal Replacement Refer to -3 for the following steps Pump assembly components Deuterium Lamp Replacement Waste port mounting plate Operating Problems TroubleshootingCommon Problems Inaccessible Fuses Circuit breaker location Teledyne Isco, Inc Service DepartmentHow to Ship Returns Appendix a Method Templates Dard.tpl Modifying Template General Editing RulesOther BlocksBlockn Appendix B Fluid Path Diagrams Fluid-type BlocksBioOptix Appendix B Fluid Path Diagrams Load-type BlocksFigure B-4 Fluid path for Load-type blocks Manual injections Plumbing Diagram Pump FlushFigure B-7 BioOptix 10 Plumbing Diagram BioOptix Appendix B Fluid Path Diagrams Appendix C Replacement Parts Lists BioOptix Appendix C Replacement Parts Lists Detector Assembly 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Column Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Pump Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Declaration of Conformity Page Teledyne Isco One Year Limited Factory Service Warranty