BioOptix™ 10

Section 2 Installation

5.Connect the AC power cord to the back of the Foxy 200. Connect the other end of the power cord to an outlet meet- ing the electrical requirements listed on the serial number label.

6.Connect the “Y” Serial Cable end labeled Second Instru- ment Foxy to the RS-232-Cport on the back of the Foxy 200.

7.Place the Foxy 200 to the right of the BioOptix 10 (Figure 2-1). Ensure that the full motion of the fraction collector arm is not obstructed, especially near column positions 9 and 10 on the BioOptix 10.

2.2.5 Fluid Connections Fluid line connections include Buffers, Samples, Waste, Pump Purge, and Foxy. Tubing for the fluid connections is located in the accessory kit, P/N 60-5209-002. The kit includes enough tubing to complete a standard installation. Standard tubing lengths are noted in the following instructions for each connection. This kit also includes a tubing cutter.

Except as noted on the Foxy connections, use the following steps for tubing connections:

1.Loosen the nut on the compression fitting.

2.Insert the tubing fully into the fitting.

3.Finger-tighten the nut.

4.Using a wrench, tighten the fitting an additional full turn.

Buffer A – These inlet ports are located on the left side of the BioOptix 10. The ferrules on these inlet ports accept the 0.125" OD tubing, P/N 023-0503-02, found in the accessories kit.

Note

Larger or smaller outside diameter tubing will require different ferrules.

Buffer A solutions may be unique to each channel, or a single solution may be routed to all ten channels.

If unique buffers will be used on each channel, cut the tubing to length (standard length is 60" or 152 cm) and route the tubing to the buffer solution containers.

If a common buffer solution will be used for all ten channels, install the Buffer Assembly (P/N

60-5204-064), found the accessory kit. Figure 2-14 illus- trates the Buffer Assembly parts. Figure 2-15 shows the buffer assembly installed on the BioOptix 10.

2-10

Image 28
Contents BioOptix Page Foreword Page General Warnings Application, take heed in the following general warningsDamage Hazard Symbols Symboles de sécuritéBioOptix 10 Safety Table of Contents BioOptixSystem Maintenance Appendix B Fluid Path DiagramsAppendix C Replacement Parts Lists Appendix a Method TemplatesList of Figures List of Tables Introduction Product OverviewBioOptix 10 Specifications Maximum Altitude 2000 metersBioOptix 10 Front Panel Controls System Controls and ConnectorsBioOptix 10 Front Panel Controls BioOptix 10 Left-side Connections BioOptix 10 Left-side Controls and ConnectionsBioOptix 10 Right-side Connections BioOptix 10 Right-side ConnectionsBioOptix 10 Rear Panel Controls and Connections BioOptix 10 Rear Panel Controls and ConnectionsInstallation BioOptix 10 InstallationUnpacking the BioOptix 10 System SpaceAssembling ElectricalBioOptix Accessing pump and column module internal connections Routing the Sample Out lines column module P110 Connection on BD3 AC power connection from pump module 10 Liquid connections pump module shown 11 Aligning the fasteners Ceeding with the arm assembly instructions FoxyBioOptix Installation Common Buffer Inlet Mounting Block Manifold Tubing Color chart Channel Numbers Color Part Number16 Foxy 200 tubing connections Windows PC Minimum Requirements ConfiguringBioOptix 10 Software BioOptix 10 System Preparation 18 Column bypass tubing 20 BioOptix 10 Manual Hardware Control window BioOptix Installation Operation Sample Preparation Column InstallationAdapters Gradients BioOptix 10 SoftwareOverview Methods Defining a Gradient To define a gradientEquilibrate tab settings Load Settings This block loads the sample onto the column Load tab settingsWash Settings Flush Settings ExperimentsSetting up Running ExperimentOperation Buttons FileZoom and pan OperationPost-experiment Operation Pump FlushSaving Run Data PreviewAutoReports BioOptix 10 Operation System Maintenance Vent damageExterior Cleaning Piston Seal Replacement Refer to -3 for the following steps Pump assembly components Deuterium Lamp Replacement Waste port mounting plate Common Problems TroubleshootingOperating Problems Inaccessible Fuses Circuit breaker location How to Ship Returns Service DepartmentTeledyne Isco, Inc Appendix a Method Templates Dard.tpl Modifying Template General Editing RulesBlockn BlocksOther Appendix B Fluid Path Diagrams Fluid-type BlocksBioOptix Appendix B Fluid Path Diagrams Load-type BlocksFigure B-4 Fluid path for Load-type blocks Manual injections Plumbing Diagram Pump FlushFigure B-7 BioOptix 10 Plumbing Diagram BioOptix Appendix B Fluid Path Diagrams Appendix C Replacement Parts Lists BioOptix Appendix C Replacement Parts Lists Detector Assembly 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Column Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Pump Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Declaration of Conformity Page Teledyne Isco One Year Limited Factory Service Warranty