Teledyne 10 manual Refer to -3 for the following steps

Page 53

BioOptix™ 10

Section 4 System Maintenance

2

4

3

1

Figure 4-2 Access the pump assembly

Note

Refer to Figure 4-3 for the following steps.

9.Remove the bottom retainer plate.

10.On the top of the pump assembly, remove the two 10-323/8" pan head screws and lock washers.

11.Lift the piston plug and top O-ring from the glass cylinder.

12.Remove the glass cylinder and bottom O-ring. This exposes the push rod and seal.

13.The push rod shaft has two slots for a wrench. Hold the push rod at the slots with a 3/16" wrench and remove the #6 screw and washers from the bottom of the push rod.

14.Lift the push rod out of the assembly.

15.Using care to avoid damaging the push rod, cut the seal with a sharp blade. Then, pull the seal off the push rod.

16.Place the replacement seal upside-down on a firm, flat sur- face (Figure 4-4).

4-3

Image 53
Contents BioOptix Page Foreword Page Application, take heed in the following general warnings General WarningsDamage Symboles de sécurité Hazard SymbolsBioOptix 10 Safety BioOptix Table of ContentsAppendix B Fluid Path Diagrams System MaintenanceAppendix C Replacement Parts Lists Appendix a Method TemplatesList of Figures List of Tables Product Overview IntroductionMaximum Altitude 2000 meters BioOptix 10 SpecificationsBioOptix 10 Front Panel Controls System Controls and ConnectorsBioOptix 10 Front Panel Controls BioOptix 10 Left-side Controls and Connections BioOptix 10 Left-side ConnectionsBioOptix 10 Right-side Connections BioOptix 10 Right-side ConnectionsBioOptix 10 Rear Panel Controls and Connections BioOptix 10 Rear Panel Controls and ConnectionsBioOptix 10 Installation InstallationUnpacking the BioOptix 10 System SpaceBioOptix ElectricalAssembling Accessing pump and column module internal connections Routing the Sample Out lines column module P110 Connection on BD3 AC power connection from pump module 10 Liquid connections pump module shown 11 Aligning the fasteners Foxy Ceeding with the arm assembly instructionsBioOptix Installation Common Buffer Inlet Mounting Block Manifold Channel Numbers Color Part Number Tubing Color chart16 Foxy 200 tubing connections BioOptix 10 Software ConfiguringWindows PC Minimum Requirements BioOptix 10 System Preparation 18 Column bypass tubing 20 BioOptix 10 Manual Hardware Control window BioOptix Installation Adapters Sample Preparation Column InstallationOperation Overview BioOptix 10 SoftwareGradients Defining a Gradient To define a gradient MethodsEquilibrate tab settings Load tab settings Load Settings This block loads the sample onto the columnWash Settings Experiments Flush SettingsSetting up Experiment RunningOperation Buttons FileOperation Zoom and panPump Flush Post-experiment OperationPreview Saving Run DataAutoReports BioOptix 10 Operation Vent damage System MaintenanceExterior Cleaning Piston Seal Replacement Refer to -3 for the following steps Pump assembly components Deuterium Lamp Replacement Waste port mounting plate Operating Problems TroubleshootingCommon Problems Inaccessible Fuses Circuit breaker location Teledyne Isco, Inc Service DepartmentHow to Ship Returns Appendix a Method Templates Dard.tpl General Editing Rules Modifying TemplateOther BlocksBlockn Fluid-type Blocks Appendix B Fluid Path DiagramsLoad-type Blocks BioOptix Appendix B Fluid Path DiagramsFigure B-4 Fluid path for Load-type blocks Manual injections Pump Flush Plumbing DiagramFigure B-7 BioOptix 10 Plumbing Diagram BioOptix Appendix B Fluid Path Diagrams Appendix C Replacement Parts Lists BioOptix Appendix C Replacement Parts Lists Detector Assembly 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Column Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Pump Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Declaration of Conformity Page Teledyne Isco One Year Limited Factory Service Warranty