Teledyne manual BioOptix 10 Manual Hardware Control window

Page 35

BioOptix™ 10

Section 2 Installation

Figure 2-20 BioOptix 10 Manual Hardware Control window

9.In the Columns section of this window, check each channel number that should be primed. Configure the remaining settings to match those shown in Figure 2-20.

10.Click the Start Pump button. Observe the following for each enabled channel:

Buffer B solution movement should be visible at the inlet ports on the left side of the BioOptix 10.

After passing through the internal pump and valves, Buffer B solution should be visible at the Pump Purge port.

11.Continue to manually pump Buffer B solution through the system and inspect the tubing connections.

If any connections are leaking, use a wrench to tighten the fitting 1/4 turn, then check again for leaks. Repeat as neces- sary.

If no leaks are found, proceed with step 12.

12.Click the Stop Pump button.

13.Change the %B setting to 0 percent.

14.Click the Start Pump button. Monitor the flow of Buffer A through the system. Correct leaks as needed.

15.Click the Stop Pump button.

16.Click the Done button.

The system is ready for operation (Section 3).

2-17

Image 35
Contents BioOptix Page Foreword Page Application, take heed in the following general warnings General WarningsDamage Symboles de sécurité Hazard SymbolsBioOptix 10 Safety BioOptix Table of ContentsAppendix a Method Templates System MaintenanceAppendix B Fluid Path Diagrams Appendix C Replacement Parts ListsList of Figures List of Tables Product Overview IntroductionMaximum Altitude 2000 meters BioOptix 10 SpecificationsBioOptix 10 Front Panel Controls System Controls and ConnectorsBioOptix 10 Front Panel Controls BioOptix 10 Left-side Controls and Connections BioOptix 10 Left-side ConnectionsBioOptix 10 Right-side Connections BioOptix 10 Right-side ConnectionsBioOptix 10 Rear Panel Controls and Connections BioOptix 10 Rear Panel Controls and ConnectionsSpace InstallationBioOptix 10 Installation Unpacking the BioOptix 10 SystemBioOptix ElectricalAssembling Accessing pump and column module internal connections Routing the Sample Out lines column module P110 Connection on BD3 AC power connection from pump module 10 Liquid connections pump module shown 11 Aligning the fasteners Foxy Ceeding with the arm assembly instructionsBioOptix Installation Common Buffer Inlet Mounting Block Manifold Channel Numbers Color Part Number Tubing Color chart16 Foxy 200 tubing connections BioOptix 10 Software ConfiguringWindows PC Minimum Requirements BioOptix 10 System Preparation 18 Column bypass tubing 20 BioOptix 10 Manual Hardware Control window BioOptix Installation Adapters Sample Preparation Column InstallationOperation Overview BioOptix 10 SoftwareGradients Defining a Gradient To define a gradient MethodsEquilibrate tab settings Load tab settings Load Settings This block loads the sample onto the columnWash Settings Experiments Flush SettingsSetting up File RunningExperiment Operation ButtonsOperation Zoom and panPump Flush Post-experiment OperationPreview Saving Run DataAutoReports BioOptix 10 Operation Vent damage System MaintenanceExterior Cleaning Piston Seal Replacement Refer to -3 for the following steps Pump assembly components Deuterium Lamp Replacement Waste port mounting plate Operating Problems TroubleshootingCommon Problems Inaccessible Fuses Circuit breaker location Teledyne Isco, Inc Service DepartmentHow to Ship Returns Appendix a Method Templates Dard.tpl General Editing Rules Modifying TemplateOther BlocksBlockn Fluid-type Blocks Appendix B Fluid Path DiagramsLoad-type Blocks BioOptix Appendix B Fluid Path DiagramsFigure B-4 Fluid path for Load-type blocks Manual injections Pump Flush Plumbing DiagramFigure B-7 BioOptix 10 Plumbing Diagram BioOptix Appendix B Fluid Path Diagrams Appendix C Replacement Parts Lists BioOptix Appendix C Replacement Parts Lists Detector Assembly 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Column Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Pump Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Declaration of Conformity Page Teledyne Isco One Year Limited Factory Service Warranty