Teledyne 10 manual Exterior Cleaning Piston Seal Replacement

Page 52

BioOptix™ 10

Section 4 System Maintenance

4.2 Exterior Cleaning

4.3Piston Seal Replacement

The BioOptix 10 system pumps the wash fluid, flushing the Buffer In lines, gradient formation components, pump, and purge lines. (See Figure B-6 in Appendix B.)

For general cleaning of the enclosure, use distilled water with a mild detergent. Use isopropyl alcohol for tougher stains. Avoid rubbing vigorously or using aggressive solvents like acetone on printed areas.

CAUTION

Do not immerse the instrument in a water bath. The instrument is not watertight and this action could damage the internal electronics.

Should pump performance degrade, the piston seals should be replaced.

To replace the piston seals:

WARNING

Risk of electric shock. Disconnect power before servicing. Only trained personnel may open the case enclosure.

WARNING

Earth ground bonding connector. Do not remove or disconnect.

1.Remove the rear cover (item 1, Figure 4-2) by unfastening the nine screws located on the top and back of the BioOptix 10.

2.Unfasten the bottom 2 screws and rotate the valve assem- bly up. Slide the slide lock over to catch and hold the valve assembly.

3.Remove the upper valve assembly (item 2, Figure 4-2) by unfastening the four screws. After removing, the valve assembly can be rotated upward and placed on top of the system with the cables and tubing still attached.

4.Remove the drip tray (item 3, Figure 4-2).

5.Remove the four screws (item 4, Figure 4-2) that secure the pump assembly to the side panel.

6.Remove the pan head screw from the top mounting tab (Figure 4-3).

7.Remove the 7/64" Allen head screw from the bottom mount- ing tab.

8.Lift the pump assembly up and out of the case.

Note

Do not disconnect tubing or wires. They are long enough to allow access to all sides of the pump assembly.

4-2

Image 52
Contents BioOptix Page Foreword Page General Warnings Application, take heed in the following general warningsDamage Hazard Symbols Symboles de sécuritéBioOptix 10 Safety Table of Contents BioOptixSystem Maintenance Appendix B Fluid Path DiagramsAppendix C Replacement Parts Lists Appendix a Method TemplatesList of Figures List of Tables Introduction Product OverviewBioOptix 10 Specifications Maximum Altitude 2000 metersBioOptix 10 Front Panel Controls System Controls and ConnectorsBioOptix 10 Front Panel Controls BioOptix 10 Left-side Connections BioOptix 10 Left-side Controls and ConnectionsBioOptix 10 Right-side Connections BioOptix 10 Right-side ConnectionsBioOptix 10 Rear Panel Controls and Connections BioOptix 10 Rear Panel Controls and ConnectionsInstallation BioOptix 10 InstallationUnpacking the BioOptix 10 System SpaceAssembling ElectricalBioOptix Accessing pump and column module internal connections Routing the Sample Out lines column module P110 Connection on BD3 AC power connection from pump module 10 Liquid connections pump module shown 11 Aligning the fasteners Ceeding with the arm assembly instructions FoxyBioOptix Installation Common Buffer Inlet Mounting Block Manifold Tubing Color chart Channel Numbers Color Part Number16 Foxy 200 tubing connections Windows PC Minimum Requirements ConfiguringBioOptix 10 Software BioOptix 10 System Preparation 18 Column bypass tubing 20 BioOptix 10 Manual Hardware Control window BioOptix Installation Operation Sample Preparation Column InstallationAdapters Gradients BioOptix 10 SoftwareOverview Methods Defining a Gradient To define a gradientEquilibrate tab settings Load Settings This block loads the sample onto the column Load tab settingsWash Settings Flush Settings ExperimentsSetting up Running ExperimentOperation Buttons FileZoom and pan OperationPost-experiment Operation Pump FlushSaving Run Data PreviewAutoReports BioOptix 10 Operation System Maintenance Vent damageExterior Cleaning Piston Seal Replacement Refer to -3 for the following steps Pump assembly components Deuterium Lamp Replacement Waste port mounting plate Common Problems TroubleshootingOperating Problems Inaccessible Fuses Circuit breaker location How to Ship Returns Service DepartmentTeledyne Isco, Inc Appendix a Method Templates Dard.tpl Modifying Template General Editing RulesBlockn BlocksOther Appendix B Fluid Path Diagrams Fluid-type BlocksBioOptix Appendix B Fluid Path Diagrams Load-type BlocksFigure B-4 Fluid path for Load-type blocks Manual injections Plumbing Diagram Pump FlushFigure B-7 BioOptix 10 Plumbing Diagram BioOptix Appendix B Fluid Path Diagrams Appendix C Replacement Parts Lists BioOptix Appendix C Replacement Parts Lists Detector Assembly 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Column Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Pump Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Declaration of Conformity Page Teledyne Isco One Year Limited Factory Service Warranty