Teledyne 10 manual Tubing Color chart, Channel Numbers Color Part Number

Page 30

BioOptix™ 10

Section 2 Installation

Buffer B – To form gradient buffers during purification, connect Buffer B solution(s) to these ten inlet ports. Perform the steps described in Buffer A above to complete these fluid connections.

Sample In – The ferrules on these inlet ports accept the 0.063" OD tubing. The accessory kit contains Teflon tubing of this size in five different colors. Cut the tubing to length (standard length is 36" or 91 cm), insert one end into a Sample In port, and route the other end to the sample container.

Note

Although not critical for operation, installing the colored tubing according the chart below may be helpful should there be a later need to trace the lines or monitor system performance.

Table 2-1 Tubing Color chart

Channel Numbers

Color

Part Number

1, 6

Blue

023-0505-06

 

 

 

2, 7

Yellow

023-0505-04

 

 

 

3, 8

Orange

023-0505-03

 

 

 

4, 9

Green

023-0505-05

 

 

 

5, 10

Clear

023-0505-09

 

 

 

Waste Out – The ferrules on these inlet ports accept the 0.063" OD tubing listed in Table 2-1, found in the accessories kit. Cut the tubing to length (standard length is 36" or 91 cm). Then, while observing the color-coded channel numbers, insert one end into a Waste Out port, and route to a suitable collection point.

If there might be a need to recover waste fluid (e.g. undetected purified sample), a collection container for each channel is recom- mended. Isolating waste fluid for each channel can simplify sample recovery.

Pump Purge – This port is used when the pumps clears any fluid remaining in the pump. This fluid is cleared before a run is started or during system maintenance. Connect a length of 0.250" OD Teflon tubing (P/N 023-0508-01) and use it to route the pump purge fluid to a suitable collection container. The standard length of tubing is 60" or 152 cm.

Column Positions – For initial operation, and when channels are unused during operation, tubing must be inserted between the Column Inlet and Outlet ports. Use the column bypass tube (P/N 60-5203-170) or pieces of clear 0.063" ID tubing (P/N 023-0505-09) from the bulk roll in accessory kit.

Foxy – The two bundles of tubing on the right side of the BioOptix 10 connect to the Foxy 200 Fraction Collector. Each color-coded tube is labeled with its channel number and finished with a threaded fitting.

2-12

Image 30
Contents BioOptix Page Foreword Page General Warnings Application, take heed in the following general warningsDamage Hazard Symbols Symboles de sécuritéBioOptix 10 Safety Table of Contents BioOptixAppendix C Replacement Parts Lists System MaintenanceAppendix B Fluid Path Diagrams Appendix a Method TemplatesList of Figures List of Tables Introduction Product OverviewBioOptix 10 Specifications Maximum Altitude 2000 metersSystem Controls and Connectors BioOptix 10 Front Panel ControlsBioOptix 10 Front Panel Controls BioOptix 10 Left-side Connections BioOptix 10 Left-side Controls and ConnectionsBioOptix 10 Right-side Connections BioOptix 10 Right-side ConnectionsBioOptix 10 Rear Panel Controls and Connections BioOptix 10 Rear Panel Controls and ConnectionsUnpacking the BioOptix 10 System InstallationBioOptix 10 Installation SpaceElectrical AssemblingBioOptix Accessing pump and column module internal connections Routing the Sample Out lines column module P110 Connection on BD3 AC power connection from pump module 10 Liquid connections pump module shown 11 Aligning the fasteners Ceeding with the arm assembly instructions FoxyBioOptix Installation Common Buffer Inlet Mounting Block Manifold Tubing Color chart Channel Numbers Color Part Number16 Foxy 200 tubing connections Configuring Windows PC Minimum RequirementsBioOptix 10 Software BioOptix 10 System Preparation 18 Column bypass tubing 20 BioOptix 10 Manual Hardware Control window BioOptix Installation Sample Preparation Column Installation OperationAdapters BioOptix 10 Software GradientsOverview Methods Defining a Gradient To define a gradientEquilibrate tab settings Load Settings This block loads the sample onto the column Load tab settingsWash Settings Flush Settings ExperimentsSetting up Operation Buttons RunningExperiment FileZoom and pan OperationPost-experiment Operation Pump FlushSaving Run Data PreviewAutoReports BioOptix 10 Operation System Maintenance Vent damageExterior Cleaning Piston Seal Replacement Refer to -3 for the following steps Pump assembly components Deuterium Lamp Replacement Waste port mounting plate Troubleshooting Common ProblemsOperating Problems Inaccessible Fuses Circuit breaker location Service Department How to Ship ReturnsTeledyne Isco, Inc Appendix a Method Templates Dard.tpl Modifying Template General Editing RulesBlocks BlocknOther Appendix B Fluid Path Diagrams Fluid-type BlocksBioOptix Appendix B Fluid Path Diagrams Load-type BlocksFigure B-4 Fluid path for Load-type blocks Manual injections Plumbing Diagram Pump FlushFigure B-7 BioOptix 10 Plumbing Diagram BioOptix Appendix B Fluid Path Diagrams Appendix C Replacement Parts Lists BioOptix Appendix C Replacement Parts Lists Detector Assembly 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Column Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Pump Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Declaration of Conformity Page Teledyne Isco One Year Limited Factory Service Warranty