Teledyne 10 manual Deuterium Lamp Replacement

Page 55

BioOptix™ 10 Section 4 System Maintenance

4.4Deuterium Lamp Replacement

17.Hold the push rod upside-down and insert it into the open- ing on the seal. Apply downward force until the seal snaps onto the push rod.

18.Refer to Figure 4-3 and reassemble the pump channel using new O-rings at the top and bottom of the glass cylin- der.

19.Repeat steps 10 through 18 for the remaining channels.

Note

Channel six is located below the inductor. To remove the glass cylinder on this channel, first remove the inductor by unfasten- ing the #8 pan head screw and internal lock washer.

20.After replacing the seals and O-rings on all channels, place the pump back into the case. Then, secure the pump:

a.Insert and tighten the pan head screw with lock washer into the top mounting tab (Figure 4-3).

b.Insert and tighten the Allen head screw with lock washer into the bottom mounting tab.

c.Insert and tighten the four screws with lock washer through the side of the case (item 4, Figure 4-2).

21.Place the drip tray (item 3, Figure 4-2) back into the sys- tem.

22.Return the upper valve assembly (item 2, Figure 4-2) to its original location and secure with the mounting hardware.

23.Place the back cover (item 1, Figure 4-2) on the BioOptix 10 and secure it with the mounting hardware.

Note

The seal replacement procedure is complete. Perform a pump flush (see section 4.1) to prime the system.

If peak amplitude of a standard sample (see section 4.5, Trouble- shooting) degrades over time, or if the baseline is noisy on all channels, the deuterium lamp may need to be replaced.

WARNING

Risk of electric shock. Disconnect power before servicing. Only trained personnel may open the case enclosure.

WARNING

Risk of damage to eyesight and skin. Wear ultraviolet (UV) light eye protection and protect skin from exposure if the deuterium lamp is operating.

4-5

Image 55
Contents BioOptix Page Foreword Page Application, take heed in the following general warnings General WarningsDamage Symboles de sécurité Hazard SymbolsBioOptix 10 Safety BioOptix Table of ContentsAppendix a Method Templates System MaintenanceAppendix B Fluid Path Diagrams Appendix C Replacement Parts ListsList of Figures List of Tables Product Overview IntroductionMaximum Altitude 2000 meters BioOptix 10 SpecificationsBioOptix 10 Front Panel Controls System Controls and ConnectorsBioOptix 10 Front Panel Controls BioOptix 10 Left-side Controls and Connections BioOptix 10 Left-side ConnectionsBioOptix 10 Right-side Connections BioOptix 10 Right-side ConnectionsBioOptix 10 Rear Panel Controls and Connections BioOptix 10 Rear Panel Controls and ConnectionsSpace InstallationBioOptix 10 Installation Unpacking the BioOptix 10 SystemAssembling ElectricalBioOptix Accessing pump and column module internal connections Routing the Sample Out lines column module P110 Connection on BD3 AC power connection from pump module 10 Liquid connections pump module shown 11 Aligning the fasteners Foxy Ceeding with the arm assembly instructionsBioOptix Installation Common Buffer Inlet Mounting Block Manifold Channel Numbers Color Part Number Tubing Color chart16 Foxy 200 tubing connections Windows PC Minimum Requirements ConfiguringBioOptix 10 Software BioOptix 10 System Preparation 18 Column bypass tubing 20 BioOptix 10 Manual Hardware Control window BioOptix Installation Operation Sample Preparation Column InstallationAdapters Gradients BioOptix 10 SoftwareOverview Defining a Gradient To define a gradient MethodsEquilibrate tab settings Load tab settings Load Settings This block loads the sample onto the columnWash Settings Experiments Flush SettingsSetting up File RunningExperiment Operation ButtonsOperation Zoom and panPump Flush Post-experiment OperationPreview Saving Run DataAutoReports BioOptix 10 Operation Vent damage System MaintenanceExterior Cleaning Piston Seal Replacement Refer to -3 for the following steps Pump assembly components Deuterium Lamp Replacement Waste port mounting plate Common Problems TroubleshootingOperating Problems Inaccessible Fuses Circuit breaker location How to Ship Returns Service DepartmentTeledyne Isco, Inc Appendix a Method Templates Dard.tpl General Editing Rules Modifying TemplateBlockn BlocksOther Fluid-type Blocks Appendix B Fluid Path DiagramsLoad-type Blocks BioOptix Appendix B Fluid Path DiagramsFigure B-4 Fluid path for Load-type blocks Manual injections Pump Flush Plumbing DiagramFigure B-7 BioOptix 10 Plumbing Diagram BioOptix Appendix B Fluid Path Diagrams Appendix C Replacement Parts Lists BioOptix Appendix C Replacement Parts Lists Detector Assembly 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Column Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Pump Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Declaration of Conformity Page Teledyne Isco One Year Limited Factory Service Warranty