Teledyne 10 manual Methods, Defining a Gradient To define a gradient

Page 39

BioOptix™ 10

Section 3 Operation

Note

The percent B value is the amount of Buffer B solution mixed with Buffer A. A value of zero percent B (0 %B) pumps only Buffer A. Conversely, a value of 100 %B pumps only Buffer B.

All other rows in the table define the %B that will be reached after pumping the listed volume. The system forms a linear gra- dient between inflection points. This can be observed in Figure 3-1. After pumping the initial 5 mL of 0 %B, a linear gradient is formed, reaching 80 %B after pumping an additional 20 mL.

A gradient file may contain as many as 9 inflection points.

3.4.1 Defining a Gradient

To define a gradient:

Using the table, enter the %B value and the pumped Volume in mL over which that value will be reached. Or,

Click in the plot area to add inflection points to the right of the plot. After adding a point you can click and drag the inflection point to change it.

Note that whenever the table or plot area is edited, both are updated.

The buttons below the table modify the table of inflection points.

Insert Row – adds a row above the selected row.

Delete Row – removes the selected row from the table.

Add Row – adds a row to the end of the table.

Sharp changes can be made in the gradient by entering a row with 0 mL as the pumped volume, followed by another row with the new %B and volume. Pairing rows such as these can be used to form step gradients.

3.4.2Saving the Gradient To be used in method and experiment files, the gradient file must be saved. Save the gradient by selecting the File>Save menu command. Use a descriptive file name so that it can be located later.

3.5 Methods

A method file can be created by:

 

• selecting the File>New>Method menu command,

 

• pressing the Ctrl+M keyboard buttons,

 

• or by clicking the M toolbar button.

 

A new method file displays a series of tabs. Each tab contains the

 

settings for each block of the method. A block is a segment of the

 

method that pumps buffer solution, loads a sample, or forms a

 

gradient.

About Method Templates

When creating a new method, the BioOptix 10 software con-

 

figures the blocks of the method file according to a template

 

file—standard.tpl. This template file can be modified to suit your

 

applications, allowing up to ten blocks or tabs. Refer to Appendix

 

A for instructions.

3-3

Image 39
Contents BioOptix Page Foreword Page Application, take heed in the following general warnings General WarningsDamage Symboles de sécurité Hazard SymbolsBioOptix 10 Safety BioOptix Table of ContentsAppendix a Method Templates System MaintenanceAppendix B Fluid Path Diagrams Appendix C Replacement Parts ListsList of Figures List of Tables Product Overview IntroductionMaximum Altitude 2000 meters BioOptix 10 SpecificationsSystem Controls and Connectors BioOptix 10 Front Panel ControlsBioOptix 10 Front Panel Controls BioOptix 10 Left-side Controls and Connections BioOptix 10 Left-side ConnectionsBioOptix 10 Right-side Connections BioOptix 10 Right-side ConnectionsBioOptix 10 Rear Panel Controls and Connections BioOptix 10 Rear Panel Controls and ConnectionsSpace InstallationBioOptix 10 Installation Unpacking the BioOptix 10 SystemElectrical AssemblingBioOptix Accessing pump and column module internal connections Routing the Sample Out lines column module P110 Connection on BD3 AC power connection from pump module 10 Liquid connections pump module shown 11 Aligning the fasteners Foxy Ceeding with the arm assembly instructionsBioOptix Installation Common Buffer Inlet Mounting Block Manifold Channel Numbers Color Part Number Tubing Color chart16 Foxy 200 tubing connections Configuring Windows PC Minimum RequirementsBioOptix 10 Software BioOptix 10 System Preparation 18 Column bypass tubing 20 BioOptix 10 Manual Hardware Control window BioOptix Installation Sample Preparation Column Installation OperationAdapters BioOptix 10 Software GradientsOverview Defining a Gradient To define a gradient MethodsEquilibrate tab settings Load tab settings Load Settings This block loads the sample onto the columnWash Settings Experiments Flush SettingsSetting up File RunningExperiment Operation ButtonsOperation Zoom and panPump Flush Post-experiment OperationPreview Saving Run DataAutoReports BioOptix 10 Operation Vent damage System MaintenanceExterior Cleaning Piston Seal Replacement Refer to -3 for the following steps Pump assembly components Deuterium Lamp Replacement Waste port mounting plate Troubleshooting Common ProblemsOperating Problems Inaccessible Fuses Circuit breaker location Service Department How to Ship ReturnsTeledyne Isco, Inc Appendix a Method Templates Dard.tpl General Editing Rules Modifying TemplateBlocks BlocknOther Fluid-type Blocks Appendix B Fluid Path DiagramsLoad-type Blocks BioOptix Appendix B Fluid Path DiagramsFigure B-4 Fluid path for Load-type blocks Manual injections Pump Flush Plumbing DiagramFigure B-7 BioOptix 10 Plumbing Diagram BioOptix Appendix B Fluid Path Diagrams Appendix C Replacement Parts Lists BioOptix Appendix C Replacement Parts Lists Detector Assembly 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Column Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Pump Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Declaration of Conformity Page Teledyne Isco One Year Limited Factory Service Warranty