BioOptix™ 10 |
| |
Section 4 | System Maintenance |
|
|
|
|
4.5.2 | Leaks | If leaks occur or if solid material is observed around a fitting, |
|
| determine the source of the leak and tighten the fitting. |
|
|
|
|
| Note |
|
| When you tighten a fitting, make sure that you do not over- |
|
| tighten it. Overtightening a fitting may lead to permanent dam- |
|
| age and require replacement of the fitting. |
4.5.3 | Electrical | 1. If your instrument stops working, check the line cord con- |
|
| nection. |
|
| 2. If the line cord is connected properly, check the fuses in the |
|
| power input module located on the BioOptix 10 rear panel. |
|
| These are time delay fuses |
|
| the mains input power voltage: |
|
| ❍ 100 to 129 VAC mains input: 2.0 A fuse |
|
| ❍ 215 to 264 VAC mains input: 1.0 A fuse. |
|
| Always replace with the same type. |
4.5.4 | Inaccessible Fuses |
|
|
|
|
|
| WARNING |
Risk of electrical shock. Disconnect the electrical power before opening the rear panel or otherwise gaining access inside the equipment. Only trained service personnel may gain access to the interior.
Internal fuses are not accessible to the user and should only be replaced by a trained service technician.
Pump Drive Board | F101 — 4.0 amp “T” type fuse. | |
Power One PS | F1 | — Not Serviceable |
Lamp Drive Board | F1 | |
Case Bottom Assy | F4 | — Not Serviceable (T2 primary). |
| F5 | — Not Serviceable (T2 secondary). |
| F6 | — Not Serviceable (T2 secondary). |
4.5.5Inaccessible Breakers The Column Module has an internal 1.2 amp circuit breaker that protects the lamp circuits and a thermal breaker to protect the deuterium lamp. These should only be reset if the lamp fails to light.
WARNING
Risk of electrical shock. Disconnect the electrical power before opening the rear panel or otherwise gaining access inside the equipment. Only trained service personnel may gain access to the interior.