Teledyne 10 manual Inaccessible Fuses

Page 58

BioOptix™ 10

 

Section 4

System Maintenance

 

 

 

 

4.5.2

Leaks

If leaks occur or if solid material is observed around a fitting,

 

 

determine the source of the leak and tighten the fitting.

 

 

 

 

 

Note

 

 

When you tighten a fitting, make sure that you do not over-

 

 

tighten it. Overtightening a fitting may lead to permanent dam-

 

 

age and require replacement of the fitting.

4.5.3

Electrical

1. If your instrument stops working, check the line cord con-

 

 

nection.

 

 

2. If the line cord is connected properly, check the fuses in the

 

 

power input module located on the BioOptix 10 rear panel.

 

 

These are time delay fuses (T-type) and rated according to

 

 

the mains input power voltage:

 

 

100 to 129 VAC mains input: 2.0 A fuse

 

 

215 to 264 VAC mains input: 1.0 A fuse.

 

 

Always replace with the same type.

4.5.4

Inaccessible Fuses

 

 

 

 

 

 

WARNING

Risk of electrical shock. Disconnect the electrical power before opening the rear panel or otherwise gaining access inside the equipment. Only trained service personnel may gain access to the interior.

Internal fuses are not accessible to the user and should only be replaced by a trained service technician.

Pump Drive Board

F101 — 4.0 amp “T” type fuse.

Power One PS

F1

— Not Serviceable

Lamp Drive Board

F1

— .375 amp “T” type fuse.

Case Bottom Assy

F4

— Not Serviceable (T2 primary).

 

F5

— Not Serviceable (T2 secondary).

 

F6

— Not Serviceable (T2 secondary).

4.5.5Inaccessible Breakers The Column Module has an internal 1.2 amp circuit breaker that protects the lamp circuits and a thermal breaker to protect the deuterium lamp. These should only be reset if the lamp fails to light.

WARNING

Risk of electrical shock. Disconnect the electrical power before opening the rear panel or otherwise gaining access inside the equipment. Only trained service personnel may gain access to the interior.

4-8

Image 58
Contents BioOptix Page Foreword Page General Warnings Application, take heed in the following general warningsDamage Hazard Symbols Symboles de sécuritéBioOptix 10 Safety Table of Contents BioOptixAppendix C Replacement Parts Lists System MaintenanceAppendix B Fluid Path Diagrams Appendix a Method TemplatesList of Figures List of Tables Introduction Product OverviewBioOptix 10 Specifications Maximum Altitude 2000 metersBioOptix 10 Front Panel Controls System Controls and ConnectorsBioOptix 10 Front Panel Controls BioOptix 10 Left-side Connections BioOptix 10 Left-side Controls and ConnectionsBioOptix 10 Right-side Connections BioOptix 10 Right-side ConnectionsBioOptix 10 Rear Panel Controls and Connections BioOptix 10 Rear Panel Controls and ConnectionsUnpacking the BioOptix 10 System InstallationBioOptix 10 Installation SpaceAssembling ElectricalBioOptix Accessing pump and column module internal connections Routing the Sample Out lines column module P110 Connection on BD3 AC power connection from pump module 10 Liquid connections pump module shown 11 Aligning the fasteners Ceeding with the arm assembly instructions FoxyBioOptix Installation Common Buffer Inlet Mounting Block Manifold Tubing Color chart Channel Numbers Color Part Number16 Foxy 200 tubing connections Windows PC Minimum Requirements ConfiguringBioOptix 10 Software BioOptix 10 System Preparation 18 Column bypass tubing 20 BioOptix 10 Manual Hardware Control window BioOptix Installation Operation Sample Preparation Column InstallationAdapters Gradients BioOptix 10 SoftwareOverview Methods Defining a Gradient To define a gradientEquilibrate tab settings Load Settings This block loads the sample onto the column Load tab settingsWash Settings Flush Settings ExperimentsSetting up Operation Buttons RunningExperiment FileZoom and pan OperationPost-experiment Operation Pump FlushSaving Run Data PreviewAutoReports BioOptix 10 Operation System Maintenance Vent damageExterior Cleaning Piston Seal Replacement Refer to -3 for the following steps Pump assembly components Deuterium Lamp Replacement Waste port mounting plate Common Problems TroubleshootingOperating Problems Inaccessible Fuses Circuit breaker location How to Ship Returns Service DepartmentTeledyne Isco, Inc Appendix a Method Templates Dard.tpl Modifying Template General Editing RulesBlockn BlocksOther Appendix B Fluid Path Diagrams Fluid-type BlocksBioOptix Appendix B Fluid Path Diagrams Load-type BlocksFigure B-4 Fluid path for Load-type blocks Manual injections Plumbing Diagram Pump FlushFigure B-7 BioOptix 10 Plumbing Diagram BioOptix Appendix B Fluid Path Diagrams Appendix C Replacement Parts Lists BioOptix Appendix C Replacement Parts Lists Detector Assembly 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Column Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Pump Module 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists 13BioOptix 10 Replacement Parts Lists Declaration of Conformity Page Teledyne Isco One Year Limited Factory Service Warranty