If oxygen is used with the ventilator, the oxygen flow must be turned off when the ventilator is not operating.

Explanation: When the ventilator is not in operation, and the oxygen flow is left on, oxygen delivered into the ventilator tubing may accumulate within the ventilator enclosure. Oxygen accumulated in the ventilator enclosure will create a risk of fire. This warning applies to most types of ventilators.

Please refrain from smoking while oxygen is in use.

About the Mask

The mask system does not contain latex, PVC or DEHP materials. If the patient has ANY reaction to any part of the mask system, discontinue use and investigate the possible causes.

ResMed has tinted the Ultra Mirage NV Full Face Mask elbow blue to identify it as a non-vented mask for use with active exhaust non-invasive ventilator systems only.

The mask has a standard female 22 mm tapered fitting (ISO5356-1) for external connection to the ventilator.

Refer to the section “Technical Specifications” of this guide for information about mask leak, dead space and resistance.

CAUTION

The mask must be cleaned and disinfected/sterilized if it is used between patients. Refer to the section “Cleaning the Mask between Patients” for details.

Parts of the Mask

See section A on the illustrations sheet.

Fitting the Mask

Note: Refer to the relevant ventilator system operating instructions for detailed information about connecting the mask.

See section B on the illustrations sheet.

The Ultra Mirage NV Full Face Mask is supplied fully assembled.

1Move the forehead support tab to Position 3. (Refer illustration B-1 for location of Positon 3.)

2Disconnect one lower headgear clip. Place the mask on the face and pull the headgear over the patient’s head (B-2). Ensure that the upper straps pass across the forehead and the lower straps below the ears.

3Reattach the headgear clip to the mask frame.

4 Gently tighten the upper headgear straps (B-3).

2

Image 6
Contents Ultra Mirage Ultra Mirage Non-vented Full Face Mask BC-6 Ultra MirageTM Intended Use About the Mask Parts of the MaskFitting the Mask Removing the Mask Gently tighten the lower headgear straps B-4Daily/After each use Disassembling for CleaningCleaning the Mask in the Home WeeklyCleaning the Mask between Patients Detergent Solution any oneReassembling the Mask DisposalTechnical Specifications Symbols Ordering InformationLimited Warranty Page Usage prévu Avertissements ET PrécautionsPropos du masque Composants du masqueMise en place du masque Remarques Retrait du masque Démontage pour le nettoyageQuotidien/Après chaque utilisation Nettoyage du masque à domicileNettoyage du masque entre les patients HebdomadaireRemontage du masque Par ex., AlconoxCaractéristiques Fuites au masque non intentionnellesEspace mort Pression de fonctionnement SymbolesNormes RésistanceInformations de commande MasqueGarantie limitée Uso indicado PrecauciónAcerca de la mascarilla Piezas de la mascarillaColocación de la mascarilla Notas Extracción de la mascarilla Desmontaje para la limpiezaDiariamente o después de cada uso Limpieza de la mascarilla en el domicilioLimpieza de la mascarilla entre pacientes SemanalmenteMontaje de la mascarilla Detergente Solución cualquieraEspecificaciones técnicas Fugas accidentales de la mascarillaEspacio muerto Normas DimensionesSímbolos Presión de funcionamientoInformación para solicitudes de envío SistemaGarantía limitada Função Advertências E PrecauçõesSobre a Máscara Peças da MáscaraColocação da Máscara Remoção da Máscara Desmontagem para LimpezaLimpeza da Máscara no Lar Limpeza da Máscara entre PacientesDiariamente / Após cada utilização Montagem da Máscara Detergente Solução qualquer uma delasEspaço Morto Especificações TécnicasFuga Acidental da Máscara Pressão de FuncionamentoInformações sobre Pedidos ResistênciaDimensões Garantia Limitada