46
ENTER COMMUNICATION OPTION MENU:
Press a1) Press a, if necessary, until CNFGprompt appears.
Press d2) Display advances to INPT Input Menu.
Press a3) Press a, if necessary, until Display advances to COMM
Communication Options Menu.
Press d4) Display advances to C.PAR Communication Parameters
Submenu.
If Communication Option is not installed, the display shows
NONE and skips to the Color Display Menu.
COMMUNICATION PARAMETERS SUBMENU:
Allows the user to adjust Serial Communications Settings of the instrument.
When connecting an instrument to a computer or other device, the
Communications Parameters must match. Generally the default settings
(as shown in Section 5) should be utilized.
Press d5) Display advances to BAUD Baud Submenu.
BAUD SUBMENU:
Press d6) Display flashes previous selection for BAUD value.
Press b7) Scroll through the available selections: 300_, 600_, 1200,
2400, 4800, 9600, 19.2K.
Press d8) Display shows STRD stored message momentarily and then
advances to PRTY only, if it was changed, otherwise press ato
advance to PRTY Parity Submenu.
PARITYSUBMENU:
Press d9) Display flashes previous selection for “Parity”.
Press b10) Scroll through the available selections: NO, ODD, EVEN.
Press d11) Display shows STRD stored message momentarily and then
advances to DATA only, if it was changed, otherwise press ato
advance to DATA Data Bit Submenu.
DATABIT SUBMENU:
Press d12) Display flashes previous selection for “Data Bit”.
Press b13) Scroll through the available selections: 7-BIT, 8-BIT.
Press d14) Display shows STRD stored message and then advances to
STOP only, if it was changed, otherwise press ato advance to
STOP Stop Bit Submenu.