48
Press d25) Display flashes previous selection for “Echo”.
Press b26) Scroll through the available selections: NO, YES.
Press d27) Display flashes STRD stored message momentarily and then
advances to STND only if it was changed, otherwise press ato
advance to STND Communication Standard Submenu.
COMMUNICATION INTERFACE STANDARD SUBMENU:
Determines whether device should be connected to an RS-232C serial port
(as is commonly used on IBM PC-compatible computers) or via an RS-485 bus
connected through appropriate RS-232/485 converter. When used in RS-485
Mode, the device must be accessed with an appropriate Address Value as
selected in the Address Submenu described later.
Press d28) Display flashes previous selection for “Standard”.
Press b29) Scroll through the available selections: 232C, 485.
Press d30) Display shows STRD stored message momentarily and then
advances to MoDE only, if it was changed, otherwise press ato
advance to MoDE Data Flow Mode Submenu.
DATAFLOW MODE SUBMENU:
Determines whether the instrument will wait for commands and data requests
from the Serial Bus or whether the instrument will send data automatically and
continuously to the Serial Bus. Devices configured for the RS-485
Communications Standard operate properly only under Command Mode.
Press d31) Display flashes previous selection for “Mode”.
Press b32) Scroll through the available selections: CMD__“Command”,
CoNT “Continuous”.
Press d33) Display shows STRD stored message momentarily and then
advances to SEPR only, if it was changed, otherwise press ato
advance to SEPR Data Separation Submenu.
DATASEPARATION CHARACTER SUBMENU:
Determines whether data sent from the device in Continuous Data Flow Mode
will be separated by spaces or by Carriage Returns.
Press d34) Display flashes previous selection for “Separation” Submenu.
Press b35) Scroll through the available selections: SPCE “Space” or
_çR_ “Carriage Return”.
Press d36) Display shows STRD stored message momentarily and then
advances to DAT.F only, if it was changed, otherwise press ato
advance to DAT.F Data Format Submenu.