3.2.7 Alarm 2

This unit is equipped with three physical outputs that can only be configured as:

a)Alarm 1, Alarm 2 & Analog Output

b)Output 1, Alarm 2 & Analog Output

Alarm must be DISABLED if Ramp is ENABLED.

Figure 3.6 Flow Chart for Alarm 2

ENTER ALARM 2 MENU:

Press a 1) Press a, if necessary, until CNFG prompt appears. Press d 2) Display advances to INPT Input Menu.

Press a 3) Press a, if necessary, until Display advances to ALR2 Alarm 2 Menu.

Press d 4) Display advances to ABSo Absolute/Deviation Submenu.

Alarm 2 menu items are identical to Alarm 1 Menu, except Alarm 2 does not have Alarm enable/disable at Power On submenu item. Modifying Alarm Settings will not reset the instrument.

29

Page 33
Image 33
a/d/s/ iSeries manual Flow Chart for Alarm, Enter Alarm 2 Menu