31
READING ADJUST SUBMENU:
Press d10) Display flashes 1st digit of previous reading adjust value.
Press b& c11) Press band cbuttons to enter a new Reading Adjust value
(-1999 to 9999).
Press d12) Display shows STRD stored message momentarily and then
advances to SP.DV Setpoint Deviation Menu.
Reading Offset Adjustallows the user to fine tune a minor error of the transducer,
however some applications may require a large offset adjust.
(Displayed Process Value = Measured Process Value ± R.ADJ).
Reading Adjust is adjustable between -1999 to 9999. For Temperature reading only, not
Process.
SETPOINT DEVIATION ENABLE/DISABLE SUBMENU:
Press d13) Display advances to Setpoint Deviation ENBL Enable or
DSBL Disable Submenu and flashes the previous selection.
Press b14) Scroll through the available selections: ENBL or DSBL.
Press d15) Display shows STRD stored message momentarily and then
advances to the CAL1 Menu.
Setpoint Deviation Submenu, if “enabled”, allows changes to Setpoint 1 to be made
automatically to Setpoint 2. This mode is very helpful if the Process Value changes often.
In Setpoint Deviation Mode, set SP2 a certain number of degrees or counts away from
SP1 - this relation remains fixed when SP1 is changed. For instance: Setting SP1=200
and SP2=20 and enabling SP.DV means that the absolute value of SP2=20. Moving SP1
to 300, the absolute value of SP2 becomes 320.
THERMOCOUPLE FIELD CALIBRATION SUBMENU:
RTD and Process are perfectly calibrated. This section is applicable to
Thermocouple (TC) calibration only.
Be sure that the TC being used to calibrate the meter is of the type
selected in the TC submenu. Place the TC in an ice-bath (or other
0°C / 32°F environment). In ambient temperature conditions: connect the
TC to the meter, apply power to the meter.
CAUTION: Do not proceed with TC calibration unless the above conditions
have been in effect for at least one hour.
Press a16) Display shows CAL1 .
Press d17) Display shows flashing 0000 .
Press a18) Display will still show flashing 0000 .
Press d19) Display shows OUT1 (meaning Calibration is complete)
* If you accidently engage the flashing 0000 (CAL° alert) simply re-press
the last button you pressed, to avoid unintentionally mis-calibrating your meter.
°°
CAUTION: Do not perform the following steps until you fully
understand this entire section.
°°