3-50
Function Reference
Normal/Delta Marker Interpolation (Marker Setup Panel)
Normal/Delta Marker Interpolation (Marker Setup Panel)
Turns the normal/delta marker interpolation on or off. When on, the normal/
delta markers will be placed at the exact wavelength or frequency value
selected if it is within the trace range. The marker will linearly inte rp olate
between two trace data points. When off, the marker will be placed on th e
nearest trace data point.
This setting enables interpolation for all four markers whe n in normal or delta
mode.
Key Path Markers > Marker Setup > Normal/Delta Marker Interpolatio n
Related Functions Active Marker, Delta Marker
Remote
Commands CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:INTerpolation
Normal Marker Units (Marker Setup Panel)
Sets the marker X-axis readout for frequency or wavelength when the i n st ru-
ment is in a non-zero span. The X-axis default units is nm (nanometers) . Thi s
setting controls only the normal marker X-axis readout and the delta ref er-
ence readout. This setting controls all four normal markers. The ba n dwidth
and delta offset markers have their own setting.
Key Path Markers > Marker Setup > Normal Marker Units
Related Functions Active Marker
Remote
Commands CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:X:READout