7-9
Specifications and Regulatory Information
Specifications

Table7-4. Dynamic Range Specifications

Table7-5. Monochromator Input Specifications

Agilent86140B, 8 6143B,
86144BaAgilent 8614 2B, 86145B,
86146Ba
a. The 86144B and 86146B specifications are for the 50 µm internal path only.
Agilent86141B an d
86140B Option 025,
86143B w/Opt 025
In 0.1 nm resolution
bandwidthb,c
b. Excluding multiple order grating response
c. With applied input fiber 9/125 µm
1250 to 1610 nm (chop mode on)
±0.5 nm, ±1 nm, ±5 nmd,e
d. Characteristic
e. Chop mode not available on the 86144B, 86146B models
–70 dB –70 dB –70 dB
1550 nm at ±0.8 nm
(±100 GHz at 1550 nm)f
f. Average of all states of polarization
60 dB 60 dB 60 dB
1550 nm at ±0.5 nmg
(±62.5 GHz at 1550 nm)
g. Characteristic for 86140B, 86141B, 86143B, 86140B with option 025, 86143B with option 025
–58 dB 58 dB –55 dB
1550 nm at ±0.4 nmg
(±50 GHz at 1550 nm) –55 dB 55 dB –52 dB
1550 nm at ±0.2 nm d
(±25 GHz at 1550 nm) –40 dB –40 dB ----
Monochromator Input Agilent8614x Ba
a. The 86144B and 86146B specifications are for the 50 µm internal path only.
Input Return Lossb,c
b. Depends on the quality of the attached connector
c. With a 9/125 µm straight connector
>35 dB >35 dB >35 dB