
7-7
Specifications and Regulatory Information
Specifications
Table7-3. Amplitude Specifications
Amplitude Agilent 8614xBa
Sensitivityb
600-750 nmc,d
750-900 nmc.d
900-1250 nmc
1250-1610 nmc
1610-1700 nme
–60 dBm
–75 dBm
–75 dBm
–90 dBm
–80 dBm
Maximum Measurement
Powerf,g
1525-1700 nm
600-1000 nm
1000-1525 nm
+15 dBm per channel,
+30 dBm total
+15 dBm per channel,
+30 dBm total
+12 dBm per channel,
+30 dBm total
Maximum Safe Power
Total safe power
Total power within any 10 nm
portion of the spectrum
+30 dBm
+23 dBm
Absolute Accuracyh
at –20 dBm, 1310 nm/1550 nm ±0.5 dB
Scale Fidelitye,i
Autorange off ± 0.05 dB
Autorange on ± 0.07 dB
Display Scale (log scale) 0.01-20 dB/DIV.
–120 to +90 dBm
Amplitude Stability
(1310 nm, 1550 nm)
1 minutej
15 minutesg±0.01 dB
±0.02 dB