3-103
Function Reference
Wavelength Offset
Mkr 1| 2
Remote
Commands (Wavelength Line Marker 1):CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]
:SRANge:LOWer
(Wavelength Line Marker 2):CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]
:SRANge:UPPer
(Sweep Limit):SENSe:WAVelength:SRANge:UPPer
Wavelength OffsetSpecifies the wavelength offset. This is an offset between the measur ed wave-
length and the displayed wavelength. You can calibrate the TLS with a wave
meter and correct for any offset. Entering a value in the wavelength offset has
no affect on the trace.
Key Path Wavelength > Wavelength Setup > Wavelength Offset
Remote
Commands SENSe:WAVelength:OFFSet
Wavelength SetupAccesses the Wavelength Setup panel:
•center wavelength step size
•user wavelength calibration date
•wavelength calibration
•wavelength offset
•wavelengths referenced in
•wavelength units
Key Path Wavelength > Wavelength Setup
Remote
Commands CALibrate:WAVelength:STATe ON
CALibrate:WAVelength:STATe OFF
SENSe:WAVelength:OFFSet
SENSe:WAVelength:CENTer:STEP:INCRement