7-8
Specifications and Regulatory Information
Specifications
Agilent 86140B, 86143B,
86144B Agilent 86142B, 86145B,
86146B
Agilent 86141B,
86140B W/Opt 025,
86143B w/Opt 025
Flatnessk
1290-1330 nm
1525-1570 nm
1525-1610 nm
1250-1610 nml
-----
±0.2 dB
-----
-----
±0.2 dB
-----
±0.2 dB
±0.7 dB
-----
±0.2 dB
-----
-----
Polarization Dependencek,m,n
1310 nm ± 0.25 dB ± 0.12 dB ----
1530 nm, 1565 nm ± 0.2 dB ± 0.05 dB ----
1600 nm ± 0.25 dB ± 0.08 dB ----
1250 to 1650 nm ± 0.3 dB ± 0.25 dB ± 0.5 dB
a. The 86144B and 86146B specifications are for the 50 µm internal path only.
b. Sensitivity is defined as signal value >6 X RMS noise value
c. Temperature range 0 to 30° C
d. Second Order
e. Temperature range 20 to 30° C
f. Resolution bandwidth setting <channel spacing
g. Characteristic
h. For resolution ≥0.1 nm
i. Excluding amplitude errors at low power levels due to noise
j. For signals within 8 dB of top of screen
k. With applied input fiber 9/125 µm
l. Absorption of light by atmospheric moisture affects flatness at 1350-1420 nm
m. For resolution ≥0.2 nm
n. Room temperature
Amplitude Agilent 8614xBa