7-10
Specifications and Regulatory Information
Specifications
Table7-6. Sweep Specifications
Table7-7. Pulse Mode Accuracy Specifications
Sweep Agilent 8614xBa
a. The 86144B and 86146B specifications are for the 50 µm internal path only.
Maximum Sweep Rateb
b. Characteristic
40 nm/56.3 ms
Maximum Sampling Rate in
Zero Spanb50 µs/trace point
Sweep Cycle Timeb
50 nm span, auto zero off
50 nm span, auto zero on
100 nm span
500 nm span
<180 ms
<340 ms
<400 ms
<650 ms
ADC Trigger Accuracyb
Jitter (distributed uniformly)
Trigger delay range <±0.5 µs
2 µs-6.5 ms
PULSE MODE ACCURACY Agilent86140B, 8 6143B,
86144Ba
a. The 86144B and 86146B specifications are for the 50 µm internal path only.
Agilent86142B , 861 45B,
86146Ba
Agilent86141B an d
86140B Option 025,
86143B w/Opt 025
Turn On ( 2 µs after rising
edge) b,c
b. Characteristic
c. Starting from dark
< ± 0.2 dB < ± 0.2 dB < ± 0.2 dB
Turn Off ( 10 µs after falling
edge)d
d. Characteristic for 86140B, 86141B, 86143B, 86144B, 86146B, 86140B w/Opt 025, 8614 3B w/Opt 025
< ± 0.2 dB <± 0.2 dB
(30 dB extinction) < ± 0.2 dB