5-20
Status Listings
OSA Warnings
50325043 desc = I/O Error
help An unexpected error occurred during the I/O operation. Please try the
operation again. If the operation involves the floppy disk drive, try a different
floppy disk. If the error persists, please make a note of the error number and
contact the nearest Agilent Technologies Instrument support center for
assistance. In the U.S., call (800) 403-0801. See the Agilent 86140B Series Users
Guide for a listing of the Agilent sales and service offices.
5044 desc = Please cycle power to synchronize system time
help = The system time clock has been set backwards. Due to internal system
requirements it is necessary to cycle power before continuing. Please turn
power off and back on again. This must be done to ensure proper system
operation.
5045 desc = Sweep limit markers too close together
help = The sweep limit function is on and the line markers are too close
together. The OSA cannot limit the sweep range to such a small value. T he
current sweep may be using a wider range than requested. Move the line
markers farther apart or turn off sweep limit to resolve this problem.
5046 desc = AutoAlign cannot find input signal at marker
help = The auto-align procedure could not find a usable input si gnal. If there is
an active marker auto-align will attempt to align at the marker wavelength
without searching for a signal peak. This error can also occur if the input signal
amplitude is very low or if the OSA has received a large mechanical shock or if
the input signal is removed after the auto-align procedure has begun. Be sure a
signal is connected to the optical input. If there is an active marker make s ure it
is placed on the input signal or turn off all markers.
5047 desc = Signal disappeared during AutoAlign
help = The auto-align procedure did not complete. The input signal disappeared
during the auto-align procedure. Make sure the input signal is connected to the
optical input and try again.
5048 desc = Wrong trace X axis units for peak to center
help = The peak-to-center function was requested. This function will only
execute if the active trace X axis has units of meters (um nm pm). Change the
active trace to one with an X axis in meters or re-measure the a ctive trace in a
non-zero span.
Table5-6. OSA Warnings (4 of 12)
Error Number Error Description [description/explanation/examples]