3-91
Function Reference
Switch Path No Auto Align
Switch Path No Auto AlignFor 86144B/86146B Filter Mode only.
Selects the 9 µm filter mode path. You should select this function if the instru-
ment has not be moved, subjected to temperature changes >2° C, turn ed off
and warmed up for at least an hour at the start of each day or to preserve pre-
vious align data.
Key Path Appl’s > Measurement Modes > Filter Mode > Switch Path Auto Align Now
Related Functions Switch Path Auto Align Now
Remote
Commands ROUTe:PATH:EXTernal
SystemAccesses menus for changing, displaying, selecting, and viewing the instru-
ment’s functions. Some of the functions available are:
•shows HW and critical errors, warnings and notices
•sets a title to the display
•displays the firmware revision
•sets printer setup
•moves the active function area
•displays the instrument state information
•changes the display setup
•changes the time and dates
•selects the power-on state
•performs a factory preset
•upgrades the firmware
•zeros the instrument
•sets the wavelength limit
•sets the TransZ 2-3 lock
•changes the auto measure defaults
•sets the GPIB address
•sets networking parameters