Agilent Technologies E4374A cfGetSenseProbeTest, cfGetSeqStep, cfGetSeqTest, Syntax, Description

Models: E4374A E4371A E4370A

1 141
Download 141 pages 7.46 Kb
Page 77
Image 77

Language Dictionary - 6

cfGetSenseProbeTest

Syntax

int cfGetSenseProbeTest(CF_HANDLE server, CF_BOOLEAN *on_off);

Description

Returns the setting of the sense probe test. The setting is either ON or OFF. The server argument can be either a handle to a group obtained by cfOpenGroup, or a handle to all cells in the instrument if no groups are defined. When this test is enabled, the instrument periodically measures the resistance of the sense probes and checks for a value that is low enough to allow accurate voltage measurements. If the probe resistance is too high and the testing is enabled, the forming sequence will be terminated for that cell.

See Also

cfSetSenseProbeTest

cfGetSeqStep

Syntax

int cfGetSeqStep(CF_HANDLE server, int step_number, CF_SEQ_OUT *out_type, float *voltage, float *current, float *time, float *reserved);

Description

Returns parameters for the given sequence step_number. The server argument can be either a handle to a group obtained by cfOpenGroup, or a handle to all cells in the instrument if no groups are defined. If the step_number is a step that has not been defined, the value returned in *out_type is CF_STEP_UNDEFINED.

cfGetSeqTest

Syntax

int cfGetSeqTest(CF_HANDLE server, CF_READP *read_pos, int *step_number, CF_SEQ_TEST *meas_test_type, float *limit, CF_TIME_TEST *time_test_type, float *time, CF_SEQ_ACTION *action);

Description

Returns the parameters of one of the sequence tests. The server argument can be either a handle to a group obtained by cfOpenGroup, or a handle to all cells in the instrument if no groups are defined.

Sequence tests for the same step are stored within the instrument in ascending order on the time parameter. This is the order in which they are returned by successive calls to cfGetSeqTest. The value pointed to by the read_pos argument controls which test is read. Read_pos points to the special value CF_READ_FIRST to read the first test in the step. After the last test is read, subsequent calls to cfGetSeqTest will return the special value CF_READ_EOF in the value pointed to by read_pos.

77

Page 77
Image 77
Agilent Technologies E4374A, E4370A, E4371A cfGetSenseProbeTest, cfGetSeqStep, cfGetSeqTest, Syntax, Description, See Also