6 - Language Dictionary
For sequence steps of type CF_CHARGE, CF_DISCHARGE, or CF_REST, the format is:
For all other sequence step types, the format is:
1 through 256 |
|
| |
1 through n, the number defined by cfSetSeqStep | |||
time | Time in seconds since the forming sequence was triggered | ||
status | A value that indicates the status of the cell | ||
| Constant | Value | Description |
| CF_CV | 1 | constant voltage mode |
| CF_CC_POS | 2 | constant current charge mode |
| CF_CC_NEG | 4 | constant current discharge mode |
| One of the following character strings: Charge, Discharge, Rest, ACR, DCR, | ||
| TaggedACR, TaggedDCR Tagged OCV, TaggedCumAH, TaggedCumWH, | ||
| ResetCumAH, ResetCumWH. |
| |
Cell voltage measurement in volts | |||
Cell current measurement in amperes | |||
The cumulative | |||
The cumulative | |||
value | The measurement or value related to the step type (either Tagged ACR, TaggedDCR, | ||
| TaggedOCV, TaggedCumWH, TaggedCumAH). |
The measure log remains in the instrument’s memory until a new forming sequence is started with the cfInitiate function.
cfReadSerial
Syntax
int cfReadSerial (CF_HANDLE server, CF_SERIAL_PORT port, int bufsize, char *buffer, int *retcount);
Description
Reads data from one of the serial ports. The port argument can be CF_PORTA or CF_PORTB. A maximum of bufsize characters will be returned in buffer. The number of characters in the buffer is returned in retcount. If there are fewer than bufsize characters available to be read, the function returns only these characters and does not wait for the buffer to fill. The instrument stores characters from the serial ports in a FIFO buffer until they are read by the controller. The function cfSerialStatus can be used to test for serial error conditions.
See Also
cfGetSerialStatus