6 - Language Dictionary

For sequence steps of type CF_CHARGE, CF_DISCHARGE, or CF_REST, the format is: cell-numberstep-number time status entry-typevolt-readingcurr-readingamp-hourswatt-hours <newline>.

For all other sequence step types, the format is:

cell-number step-number time status entry-type value <newline>

cell-number

1 through 256

 

 

step-number

1 through n, the number defined by cfSetSeqStep

time

Time in seconds since the forming sequence was triggered

status

A value that indicates the status of the cell

 

Constant

Value

Description

 

CF_CV

1

constant voltage mode

 

CF_CC_POS

2

constant current charge mode

 

CF_CC_NEG

4

constant current discharge mode

entry-type

One of the following character strings: Charge, Discharge, Rest, ACR, DCR,

 

TaggedACR, TaggedDCR Tagged OCV, TaggedCumAH, TaggedCumWH,

 

ResetCumAH, ResetCumWH.

 

volt-reading

Cell voltage measurement in volts

curr-reading

Cell current measurement in amperes

amp-hours

The cumulative ampere-hours measured from the beginning of this step-number.

watt-hours

The cumulative watt-hours measured from the beginning of this step-number.

value

The measurement or value related to the step type (either Tagged ACR, TaggedDCR,

 

TaggedOCV, TaggedCumWH, TaggedCumAH).

The measure log remains in the instrument’s memory until a new forming sequence is started with the cfInitiate function.

cfReadSerial

Syntax

int cfReadSerial (CF_HANDLE server, CF_SERIAL_PORT port, int bufsize, char *buffer, int *retcount);

Description

Reads data from one of the serial ports. The port argument can be CF_PORTA or CF_PORTB. A maximum of bufsize characters will be returned in buffer. The number of characters in the buffer is returned in retcount. If there are fewer than bufsize characters available to be read, the function returns only these characters and does not wait for the buffer to fill. The instrument stores characters from the serial ports in a FIFO buffer until they are read by the controller. The function cfSerialStatus can be used to test for serial error conditions.

See Also

cfGetSerialStatus