Agilent Technologies E4374A, E4370A, E4371A manual A - Specifications

Models: E4374A E4371A E4370A

1 141
Download 141 pages 7.46 Kb
Page 116
Image 116
A - Specifications

A - Specifications

Table A-2. Agilent E4370A/E4374A MCCD Characteristics

Parameter

ac Resistance Measurement

dc Resistance Measurement

Ah Capacity Measurement Accuracy

Wh Capacity Measurement Accuracy

Maximum Deviation of Measured Output Current from Programmed Output Current

Condition

maximum measurable maximum time/output to measure1

maximum measurable maximum time/output to measure1

%of reading + offset

%of reading + offset constant current mode

Value

1Ω

1s

1Ω

1s

±(0.01% + 1 mAh/h)

±(0.01% + 5 mWh/h)

±2 mA

Maximum Deviation of Measured Output Voltage from Programmed Output Voltage

constant current mode

±2 mV

Voltage Output Noise2

Current Output Noise2

Maximum Current Overshoot

Maximum Voltage Overshoot

Maximum Current Risetime

Minimum Programming Current

Measurement Interval

Step Time Interval

Maximum Number of Readings in Data Buffer

rms peak-to-peak

rms peak-to-peak

maximum duration of less than 5 ms

charging and discharging mode

for data logging for sequence tests for probe check/channel3

maximum minimum

30mV 100 mV

1mA 10 mA 5 % 25 mV

0.1s

25 mA

1s

1s

1s

596hours 1 s

349,504

Maximum Sequence Length

Maximum Number of Steps in Sequence

Maximum Number of Defined Groups

ac Input Line Requirements

Maximum Sense Probe Resistance3

Auxiliary bias output power

Auxiliary bias output voltage

Auxiliary bias output current

Auxiliary bias output voltage accuracy

Auxiliary bias output noise

Notes:

for mainframes with 256 channels

input voltage range input frequency range maximum input power max. current @ 100/120 Vac

max. current @ 220/240 Vac

max. @ 5 to 24V/0.42 to 2 A

max. @ 0 to 0.42A min. @ 0 to 2 A

max. @ 5 V output min. @ 24 V output

%of setting at any voltage and current

peak to peak at any voltage and current

596hours 100

8

95 Vac - 250 Vac

47 Hz - 63 Hz

300W

4 A

2 A

1kΩ 10 W

24V 5 V

2A

0.42 A

7 %

100 mV

1To measure 256 cells takes up to 5 minutes.

2At the power connector, with a bandwidth of 20Hz to 20MHz.

3To measure output probe resistance accurately there must be 50 mV between the +/power leads and the +/sense leads. To measure sense probe resistance accurately there must be 100 mV of cell voltage.

116

Page 116
Image 116
Agilent Technologies E4371A manual A - Specifications, Table A-2. Agilent E4370A/E4374A MCCD Characteristics