5020A

Users Manual

A value of 1 is returned if serial port echo (duplex) is ON. A value of 0 is returned if serial port echo (duplex) is OFF.

SYSTem:COMMunicate:SERial:FDUPlex <bool>

Sets the serial port echo (duplex) setting. Example: SYST:COMM:SER:FDUP 1

The <bool> parameter turns serial port echo (duplex) mode on (1 or ON) or off (0 or OFF). A baud rade of 57600 suspends echo (full duplex).

SYSTem:COMMunicate:SERial:FEED?

Returns the state of the serial port auto printing. Example: SYST:COMM:SER:FEED? Response: 0

A value of 1 is returned if serial port auto printing is enabled. A value of 0 if returned is serial port auto printing is disabled.

SYSTem:COMMunicate:SERial:FEED <bool>

Sets the state of the serial port auto printing. Example: SYST:COMM:SER:FEED 1

The <bool> parameter turns serial port auto printing on (1 or ON) or off (0 or OFF). The *RST command sets the serial port auto printing to OFF.

SYSTem:COMMunicate:SERial:LINefeed?

Returns the state of the serial port linefeed setting. Example: SYST:COMM:SER:LIN?

Response: 1

A value of 1 is returned if the serial port linefeed is enabled. A value of 0 is returned if serial port linefeed is disabled.

SYSTem:COMMunicate:SERial:LINefeed <bool>

Sets the state of the serial port linefeed setting.

Example: SYST:COMM:SER:LIN 1

The <bool> parameter turns serial port linefeed on (1 or ON) or off (0 or OFF).

SYSTem:COMMunicate:SERial:TIMe? [MINMAXDEF]

Returns the serial port auto printing period in seconds. Example: SYST:COMM:SER:TIM?

Response: 10

Specifying the MIN, MAX, or DEF parameter returns the minimum, maximum, or default value respectively.

SYSTem:COMMunicate:SERial:TIMe <num>MINMAXDEF

Sets to serial port sample period setting in seconds.

6-38

Page 108
Image 108