Digital Communications Interface

6

 

 

Interface Commands

 

 

 

 

 

ALARm:TEMPerature<chn>:UPPer:ENABle?

Returns the state of the high temperature alarm

 

 

 

enable for the specified channel

 

 

 

 

 

 

ALARm:TEMPerature<chn>:UPPer:ENABle <bool>

Enables or disables the high temperature alarm for the

 

 

 

specified channel

 

 

 

 

 

 

ALARm:TEMPerature<chn>:UPPer:LIMit?

Returns the high temperture alarm limit for the

 

 

 

specified channel

 

 

 

 

 

 

ALARm:TEMPerature<chn>:UPPer:LIMit

Sets the high temperture alarm limit for the specified

 

 

<num>MINMAXDEF

channel

 

 

 

 

 

 

ALARm:TIME:FIRSt?

Returns the time of first alarm event

 

 

 

 

 

 

ALARm:TIME:LAST?

Returns the time of latest alarm event

 

 

 

 

 

 

CALibrate<chn>:ALERt?

Returns the state of the calibration alert enable for the

 

 

 

specified channel

 

 

 

 

 

 

CALibrate<chn>:ALERt <bool>

Enables or disables the sensor alert enable for the

 

 

 

specified channel

 

 

 

 

 

 

CALibrate<chn>:DATe:CALibrate?

Returns the sensor calibration date for the specified

 

 

 

channel

 

 

 

 

 

 

CALibrate<chn>:DATe:CALibrate

Sets the sensor calibration date for the specified

 

 

(<year>,<month>,<day>)

channel

 

 

 

 

 

 

CALibrate<chn>:DATe:DUE?

Returns the sensor calibration due date for the

 

 

 

specified channel

 

 

 

 

 

 

CALibrate<chn>:DATe:DUE (<year>,<month>,<day>)

Sets the sensor calibration due date for the specified

 

 

 

channel

 

 

 

 

 

 

CALibrate<chn>:EXPiration?

Returns the sensor calibration expiration for the

 

 

 

specified channel

 

 

 

 

 

 

CALibrate<chn>:PARameter:LINearity<chn>?

Returns the sensor calibration linearity for the

 

 

 

specified channel (temperature or humidity)

 

 

 

 

 

 

CALibrate<chn>:PARameter:LINearity<chn> <float>

Sets the sensor calibration linearity the specified

 

 

 

channel(temperature or humidity)

 

 

 

 

 

 

CALibrate<chn>:PARameter:OFFSets<chn>?

Returns the sensor calibration offsets for the specified

 

 

 

channel (temperature or humidity)

 

 

 

 

 

 

CALibrate<chn>:PARameter:OFFSets<chn> <float>

Sets the sensor calibration offsets for the specified

 

 

 

channel (temperature or humidity)

 

 

 

 

 

 

CALibrate<chn>:PARameter:SCALe<chn>?

Returns the sensor calibration scale for the specified

 

 

 

channel (temperature or humidity)

 

 

 

 

 

 

CALibrate<chn>:PARameter:SCALe<chn> <float>

Sets the sensor calibration scale for the specified

 

 

 

channel (temperature or humidity)

 

 

 

 

 

 

CALCulate:AVERage:CLEar

Resets all running statistics

 

 

 

 

 

 

CALCulate<chn>:DEWPoint?

Returns the dew-point measurement for the specified

 

 

 

channel

 

 

 

 

 

 

CALCulate<chn>:HINDex?

Returns the heat index measurement for the specified

 

 

 

channel

 

 

 

 

 

 

CALCulate<chn>:PARameter<num>:AVERerage<nu

Returns a running statistic for the specified channel

 

 

 

 

 

6-7

Page 77
Image 77
Fluke 5020A user manual Enables or disables the high temperature alarm for