
5020A
Users Manual
ALARm:TEMPurature<chn>:RATE:LIMit? [MINMAXDEF]
Returns the temperature rate alarm limit for the specified channel in degrees (C or F) per hour.
Example: ALAR:TEMP1:RATE:LIM?
Response: 5
The TEMPerature suffix, <chn>, specifies the channel (1 or 2). The value returned is always in the current temperature units (C or F). Specifying the MIN, MAX, or DEF parameter returns the minimum, maximum, or default value respectively.
ALARm:TEMPurature<chn>:RATE:LIMit <float>MINMAXDEF
Sets the temperature rate alarm limit for the specified channel in degrees (C or F) per hour.
Example: ALAR:TEMP1:RATE:LIM 10.00
The TEMPerature suffix, <chn>, specifies the channel (1 or 2). The <float> parameter specifies the value for the temperature rate alarm limit. The <float> value is always in the current temperature units (C or F). Specifying MIN, MAX, or DEF parameter sets the temperature rate alarm limit to the minimum, maximum, or default value respectively. This command may be password protected (see Password Commands).
ALARm:TEMPurature<chn>:SENSor?
Indicates whether a temperature sensor alarm event has occurred for the specified channel.
Example: ALAR:TEMP1:SENS?
Response: 0
The TEMPerature suffix, <chn>, specifies the channel (1 or 2). A value of 1 is returned if a temperature sensor alarm event has occurred on the specified channel. A value of 0 is returned if a temperature sensor alarm event has not occurred on the specified channel.
ALARm:TEMPurature<chn>:SENSor:ENABle?
Returns the state of the temperature sensor alarm enable for the specified channel. Example: ALAR:TEMP1:SENS:ENAB?
Response: 0
The TEMPerature suffix, <chn>, specifies the channel (1 or 2). A value of 1 is returned if the temperature sensor alarm is enabled. A value of 0 is returned if the temperature sensor alarm is disabled.
ALARm:TEMPurature<chn>:SENSor:ENABle <bool>
Sets the state of the temperature sensor alarm enable for the specified channel. Example: ALAR:TEMP1:SENS:ENAB 1
The TEMPerature suffix, <chn>, specifies the channel (1 or 2). The <bool> parameter turns the temperature sensor alarm on (1 or ON) or off (0 or OFF). This command may be password protected (see Password Commands).
ALARm:TEMPurature<chn>:UPPer?
Indicates whether an upper temperature alarm event has occurred.