5020A

Users Manual

The following parameters are set when this command is issued.

Sets the measurement period to 2 seconds.

Sets the measurement averaging on.

Sets printing to the serial port

Sets the temperature units to C.

Sets the temperature resolution to 2 decimal places and the relative humidity resolution to 1 decimal place.

Stops auto logging.

Sets the statistical functions.

SYSTem:BOOT:VERSion?

Returns the boot ROM version.

Example: SYST:BOOT:VERS?

Response: 0.31

SYSTem:ERRor?

Returns a message from the system error queue. Example: SYST:ERR?

Response: 0,"No error"

The error queue may contain up to ten messages. The messages are reported in the order they occurred. Reading a message from the queue removes the message so the next message can be read. If more than ten errors occur before being read, the tenth error in the queue is -350,"Queue overflow" and all but the first nine errors are discarded. All errors cause the Error Bit (ERR, bit 2) for the Status BYTE Register to be set. Refer to Troubleshooting for the error messages.

SYSTem:CODE:VERSion?

Returns the firmware version.

Example: SYST:CODE:VERS?

Response: 1.00

SYSTem:VERSion?

Returns the SCPI version number.

Example: SYST:VERS?

Response: 1994.0

UNIT:TEMPerature?

Returns the system temperature unit.

Example: UNIT:TEMP?

Response: C

The response is either C for Celsius or F for Fahrenheit.

6-36

Page 106
Image 106
Fluke 5020A user manual SYSTemBOOTVERSion?, SYSTemERRor?, SYSTemCODEVERSion?, SYSTemVERSion?, UNITTEMPerature?