5020A
Users Manual
Caution
When data recording memory fills up, the earliest block of data is automatically discarded so recording of the most recent data can continue.
Data Record
The DATA RECORD function allows the setting of the data recording features. The functions that appear in this menu are RECORD SETTING, DATA VIEW, DATA PRINT, DATA STORAGE, and DATA CLEAR.
avu018.jpg
Figure 5-17. Data Record
Use the A D buttons to select the menu. Press the E button to select the menu. Press the F button to cancel any changes and exit to the menu.
Record Setting
The RECORD SETTING function allows you to select what is to be recorded and how often the recording occurs. This may be
To record the temperature on sensor channel 1, set REC 1/T to ON. To record the humidity on sensor channel 1, set REC 1/H to ON. To record the temperature on sensor channel 2, set REC 2/T to ON. To record the humidity on sensor channel 2, set REC 2/H to ON.
The REC PER recording period is set by selecting 1, 2, 5, 10, 15, or 30 seconds, 1, 2, 5, 10, 20, or 30 minutes, or 1 hour for the REC PER.
Note
It is recommended that REC PER be set to 5 minutes or more; otherwise, the memory may fill up quickly and redrawing graphs may be slow. Also note that graphs can only be redrawn if the measurement for the channel is being recorded, as the graph data is obtained from recorded data. The recording period affects the observable time resolution of graphs.