Fluke 5020A user manual ALARmRHUMiditychnLOWerLIMit floatMINMAXDEF, ALARmRHUMiditychnRATE?

Models: 5020A

1 142
Download 142 pages 21.97 Kb
Page 86
Image 86
ALARm:RHUMidity<chn>:LOWer:LIMit <float>MINMAXDEF

5020A

Users Manual

Response: 20.00

The RHUMidity suffix, <chn>, specifies the channel (1 or 2). Specifying the MIN, MAX, or DEF parameter returns the minimum, maximum, or default value respectively.

ALARm:RHUMidity<chn>:LOWer:LIMit <float>MINMAXDEF

Sets the lower humidity alarm limit for the specified channel in %RH. Example: ALAR:RHUM1:LOW:LIM 25.00

The RHUMidity suffix, <chn>, specifies the channel (1 or 2). The <float> parameter specifies the value for the lower humidity alarm limit. Specifying the MIN, MAX, or DEF parameter sets the lower humidity alarm limit to the minimum, maximum, or default value respectively. This command may be password protected (see Password Commands).

ALARm:RHUMidity<chn>:RATE?

Indicates whether a humidity rate alarm event has occurred for the specified channel. Example: ALAR:RHUM1:RATE?

Response: 1

The RHUMidity suffix, <chn>, specifies the channel (1 or 2). A value of 1 is returned if a humidity rate alarm event has occurred on the specified channel. A value of 0 is returned if a humidity rate alarm event has not occurred on the specified channel.

ALARm:RHUMidity<chn>:RATE:ENABle?

Returns the state of the humidity rate alarm enable for the specified channel. Example: ALAR:RHUM1:RATE:ENAB?

Response: 0

The RHUMidity suffix, <chn>, specifies the channel (1 or 2). A value of 1 is returned if the humidity rate alarm is enabled. A value of 0 is returned if the humidity rate alarm is disabled.

ALARm:RHUMidity<chn>:RATE:ENABle <bool>

Sets the state of the humidity rate alarm enable for the specified channel. Example: ALAR:RHUM1:RATE:ENAB 1

The RHUMidity suffix, <chn>, specifies the channel (1 or 2). The <bool> parameter turns the humidity rate alarm enable on (1 or ON) or off (0 or OFF). This command may be password protected (see Password Commands).

ALARm:RHUMidity<chn>:RATE:LIMit? [MINMAXDEF]

Returns the humidity rate alarm limit for the specified channel in %RH per hour. Example: ALAR:RHUM1:RATE:LIM?

Response: 5

The RHUMidity suffix, <chn>, specifies the channel (1 or 2). Specifying the MIN, MAX, or DEF parameter returns the minimum, maximum, or default value respectively.

6-16

Page 86
Image 86
Fluke 5020A user manual ALARmRHUMiditychnLOWerLIMit floatMINMAXDEF, ALARmRHUMiditychnRATE?, ALARmRHUMiditychnRATEENABle?