Fluke 5020A STATusOPERationENABle?, STATusOPERationENABle numMINMAXDEF, STATusQUEStionable?

Models: 5020A

1 142
Download 142 pages 21.97 Kb
Page 121
Image 121
STATus:OPERation:ENABle?

Digital Communications Interface 6

Commands

register by adding the binary-weighted values of each bit. The meaning of each bit, when set (1), is as follows where 0 is the least significant and 7 is the most significant:

0:no operation

Bit 0 Not used. Bit 1 Not used. Bit 2 Not used. Bit 3 Not used.

Bit 4 Measurement complete bit. Bit 5 Not used.

Bit 6 Not used. Bit 7 Not used.

Example: STAT:OPER:COND?

Response: 16

STATus:OPERation:ENABle?

This command returns the Operation Status Enable Register (see

STATus:OPERation:ENABle <num> command below).

Example: STAT:OPER:ENAB?

Response: 16

STATus:OPERation:ENABle <num>MINMAXDEF

This command sets the Operation Status Enable Register. This register determines which event bits of the Operation Status Event Register affect the Operation Status Bit of the Status Byte Register. If any bit in the Operation Status Event Register is set (1) while its corresponding mask bit in the Operation Status Enable Register is set, the Operation Status Bit in the Status Byte Register will be set. The value of the <num> parameter is a number from 0 to 15 that is the sum of the binary-weighted values of each mask bit. The Operation Status Event Register is described in Section 8.4.11.13, STATus:OPERation?. The following example causes the Operation Status Bit in the Status Byte Register to be set when measurements on both channels have been completed.

Example: STAT:OPER:ENAB 16

Response: 16

STATus:QUEStionable?

This command reads and clears the Questionable Status Event Register. The value returned indicates the condition of each of the eight bits of the register by adding the binary-weighted values of each bit. The meaning of each bit, when set (1), is as follows where 0 is the least significant and 7 is the most significant:

0:no problem

Bit 0 Not used. Bit 1 Not used. Bit 2 Not used. Bit 3 Not used.

Bit 4 Measurement Invalid bit. Bit 5 Not used.

Bit 6 Not used. Bit 7 Not used.

Example: STAT:QUES?

Response: 16

6-51

Page 121
Image 121
Fluke 5020A user manual STATusOPERationENABle?, STATusOPERationENABle numMINMAXDEF, STATusQUEStionable?