| Introduction and Specifications | 1 | |
|
| Specifications | ||
Specifications |
|
|
| |
| Temperature Range | 0 °C to 50 °C |
|
|
|
|
|
|
|
| Temperature Accuracy | 16 °C to 24 °C: ±0.125 °C (calibrated) |
| |
| (“H” Model) | 0 °C to 16 °C, 24 °C to 50 °C: ±0.5 °C (uncalibrated |
| |
|
| typical) |
| |
|
|
|
|
|
| Temperature Accuracy | 15 °C to 35 °C: ±0.25 °C (calibrated) |
| |
| (“S” Model) | 0 °C to 15 °C, 35 °C to 50 °C: ±0.5 °C (uncalibrated |
| |
|
| typical) |
| |
|
|
|
|
|
| Delta Temperature Accuracy | ±0.025 °C for ±1 °C changes within 15 °C to 35 °C |
| |
|
|
|
|
|
| Temperature Display Resolution | User selectable up to 0.001 °C (0.0 1°C recorded) |
| |
|
|
|
|
|
| RH Range | 0% to 100% RH |
| |
|
|
|
|
|
| RH Accuracy | 20% to 70% RH: ±1.5% RH (calibrated) |
| |
| (“H” Model) | 0% to 20% RH, 70% to 100% RH: ±3% RH |
| |
|
|
| ||
|
| (uncalibrated, typical) |
| |
|
|
|
|
|
| RH Accuracy | 20% to 70% RH: ±2% RH (calibrated) |
| |
| (“S” Model) | 0% to 20% RH, 70% to 100% RH: ±3% RH |
| |
|
|
| ||
|
| (uncalibrated, typical) |
| |
|
|
|
|
|
| Delta Humidity Accuracy | ±1.0% for ±5% changes within 20% to 70% RH |
| |
|
|
|
|
|
| RH Display Resolution | User selectable up to 0.01% (0.1% recorded) |
| |
|
|
|
|
|
| Inputs | Two sensors, each measuring temperature and relative |
| |
|
| humidity; each is detachable, |
| |
|
| interchangeable, with |
| |
|
| may be assigned a unique |
| |
|
|
|
|
|
| Display | 240 x 128 graphics monochrome LCD, displays |
| |
|
| temperature and humidity data graphically, numerically, |
| |
|
| and statistically; 16 |
| |
|
|
| ||
|
|
|
|
|
| Memory | 400,000 typical individual |
| |
|
| (excluding data card storage) |
| |
|
|
|
|
|
| Alarms | Visual and audio alarms for temperature, temperature |
| |
|
| rate, RH, RH rate, and fault conditions |
| |
|
|
|
|
|
| Communications |
| ||
|
|
|
|
|
| Data Card Interface | Removable data card for transferring data to a computer; |
| |
|
| data can likewise be uploaded from a data card into the |
| |
|
| 5020A for graphical and statistical display |
| |
|
|
|
|
|
| Enclosure | The 5020A may be |
| |
|
| set on a benchtop |
| |
|
|
|
|
|
| Power | 12 V DC from external |
| |
|
|
|
|
|
| Battery Backup | Standard 9V battery to allow continued measuring during |
| |
|
| power disruptions |
| |
|
|
|
|
|
| Operating Range | 0 °C to 50 °C |
| |
|
|
|
|
|