Fluke 5020A user manual Introduction and Specifications

Models: 5020A

1 142
Download 142 pages 21.97 Kb
Page 17
Image 17

 

 

Introduction and Specifications

1

 

 

Specifications

Specifications

 

 

 

 

Temperature Range

0 °C to 50 °C

 

 

 

 

 

 

 

 

Temperature Accuracy

16 °C to 24 °C: ±0.125 °C (calibrated)

 

 

(“H” Model)

0 °C to 16 °C, 24 °C to 50 °C: ±0.5 °C (uncalibrated

 

 

 

typical)

 

 

 

 

 

 

 

Temperature Accuracy

15 °C to 35 °C: ±0.25 °C (calibrated)

 

 

(“S” Model)

0 °C to 15 °C, 35 °C to 50 °C: ±0.5 °C (uncalibrated

 

 

 

typical)

 

 

 

 

 

 

 

Delta Temperature Accuracy

±0.025 °C for ±1 °C changes within 15 °C to 35 °C

 

 

 

 

 

 

 

Temperature Display Resolution

User selectable up to 0.001 °C (0.0 1°C recorded)

 

 

 

 

 

 

 

RH Range

0% to 100% RH

 

 

 

 

 

 

 

RH Accuracy

20% to 70% RH: ±1.5% RH (calibrated)

 

 

(“H” Model)

0% to 20% RH, 70% to 100% RH: ±3% RH

 

 

 

 

 

 

(uncalibrated, typical)

 

 

 

 

 

 

 

RH Accuracy

20% to 70% RH: ±2% RH (calibrated)

 

 

(“S” Model)

0% to 20% RH, 70% to 100% RH: ±3% RH

 

 

 

 

 

 

(uncalibrated, typical)

 

 

 

 

 

 

 

Delta Humidity Accuracy

±1.0% for ±5% changes within 20% to 70% RH

 

 

 

 

 

 

 

RH Display Resolution

User selectable up to 0.01% (0.1% recorded)

 

 

 

 

 

 

 

Inputs

Two sensors, each measuring temperature and relative

 

 

 

humidity; each is detachable, cable-extendable, and

 

 

 

interchangeable, with self-contained calibration; each

 

 

 

may be assigned a unique 16-character identification

 

 

 

 

 

 

 

Display

240 x 128 graphics monochrome LCD, displays

 

 

 

temperature and humidity data graphically, numerically,

 

 

 

and statistically; 16 pre-defined, user-changeable screen

 

 

 

set-ups are included

 

 

 

 

 

 

 

Memory

400,000 typical individual time-stamped readings

 

 

 

(excluding data card storage)

 

 

 

 

 

 

 

Alarms

Visual and audio alarms for temperature, temperature

 

 

 

rate, RH, RH rate, and fault conditions

 

 

 

 

 

 

 

Communications

RS-232 and IR

 

 

 

 

 

 

 

Data Card Interface

Removable data card for transferring data to a computer;

 

 

 

data can likewise be uploaded from a data card into the

 

 

 

5020A for graphical and statistical display

 

 

 

 

 

 

 

Enclosure

The 5020A may be wall-mounted (hardware included) or

 

 

 

set on a benchtop

 

 

 

 

 

 

 

Power

12 V DC from external 100-240 VAC power supply

 

 

 

 

 

 

 

Battery Backup

Standard 9V battery to allow continued measuring during

 

 

 

power disruptions

 

 

 

 

 

 

 

Operating Range

0 °C to 50 °C

 

 

 

 

 

 

1-7

Page 17
Image 17
Fluke 5020A user manual Introduction and Specifications