Fluke 5020A user manual

Models: 5020A

1 142
Download 142 pages 21.97 Kb
Page 76
Image 76

5020A Users Manual

ALARm:RHUMidity<chn>:SENSor:ENABle?

Returns the state of the humidity sensor alarm enable

 

for the specified channel

 

 

ALARm:RHUMidity<chn>:SENSor:ENABle <bool>

Enables or disables the humidity sensor alarm for the

 

specified channel

 

 

ALARm:RHUMidity<chn>:UPPer?

Returns the state of the high humidity event for the

 

specified channel

 

 

ALARm:RHUMidity<chn>:UPPer:ENABle?

Returns the state of the high humidity alarm enable for

 

the specified channel

 

 

ALARm:RHUMidity<chn>:UPPer:ENABle <bool>

Enables or disables the high humidity alarm for the

 

specified channel

 

 

ALARm:RHUMidity<chn>:UPPer:LIMit?

Returns the high humidity alarm limit for the specified

 

channel

 

 

ALARm:RHUMidity<chn>:UPPer:LIMit

Sets the high humidity alarm limit for the specified

<num>MINMAXDEF

channel

 

 

ALARm:TEMPerature<chn>:LOW?

Returns the state of the low temperature alarm for the

 

specified channel

 

 

ALARm:TEMPerature<chn>:LOW:ENABle?

Returns the state of the low temperature alarm enable

 

for the specified channel

 

 

ALARm:TEMPerature<chn>:LOW:ENABle <bool>

Enables or disables the low temperature alarm for the

 

specified channel

 

 

ALARm:TEMPerature<chn>:LOW:LIMit?

Returns the low temperature alarm limit for the

 

specified channel

 

 

ALARm:TEMPerature<chn>:LOW:LIMit

Sets the low temperature alarm limit for the specified

<num>MINMAXDEF

channel

 

 

ALARm:TEMPerature<chn>:RATE?

Returns the state of the temperature rate alarm for the

 

specified channel

 

 

ALARm:TEMPerature<chn>:RATE:ENABle?

Returns the state of the temperature rate alarm enable

 

for the specified channel

 

 

ALARm:TEMPerature<chn>:RATE:ENABle <bool>

Enables or disables the temperature rate alarm for the

 

specified channel

 

 

ALARm:TEMPerature<chn>:RATE:LIMit?

Returns the temperature rate alarm limit for the

 

specified channel

 

 

ALARm:TEMPerature<chn>:RATE:LIMit

Sets the temperature rate alarm limit for the specified

<num>MINMAXDEF

channel

 

 

ALARm:TEMPerature<chn>:SENSor?

Returns the state of the temperature sensor alarm for

 

the specified channel

 

 

ALARm:TEMPerature<chn>:SENSor:ENABle?

Returns the state of the temperature sensor alarm

 

enable for the specified channel

 

 

ALARm:TEMPerature<chn>:SENSor:ENABle <bool>

Enables or disables the temperature sensor alarm for

 

the specified channel

 

 

ALARm:TEMPerature<chn>:UPPer?

Returns the state of the high temperature alarm for the

 

specified channel

 

 

6-6

Page 76
Image 76
Fluke 5020A user manual