5020A

Users Manual

Data Commands

The data commands are used to access daily statistics or recorded measurements.

DATa:DSTatistics:ENABle?

Returns the state of the daily statistics recording enable. Example: DAT:DST:ENAB?

Response: 0

A value of 1 is returned if the daily statistics recording is enabled. A value of 0 is returned if the daily statistics recording is disabled.

DATa:DSTatistics:ENABle <bool>

Sets the state of the daily statistics recording. Example: DAT:DST:ENAB 1

The <bool> parameter turns the daily statistics recording on (1 or ON) or off (0 or OFF). This command may be password protected (see Password Commands).

DATa:DSTatistics:HOUR? [MINMAXDEF]

Returns the daily statistics recording time. Example: DAT:DST:HOUR?

Response: 12

Specifying the MIN, MAX, or DEF parameter returns the minimum, maximum, or default value respectively.

DATa:DSTatistics:HOUR <num>MINMAXDEF

Sets the state of the daily statistics recording time. Example: DAT:DST:HOUR 1

The <num> parameter specifies the hour for the daily statistics recording time. Specifying MIN, MAX, or DEF parameter sets the daily statistics recording time to the minimum, maximum, or default value respectively. This command may be password protected (see Password Commands).

DATa:DSTatistics:RENable?

Returns the state of the daily statistics reset enable. Example: DAT:DST:REN?

Response: 0

A value of 1 is returned if the daily statistics reset is enabled. A value of 0 is returned if the daily statistics reset is disabled.

DATa:DSTatistics:RENable <bool>

Sets the state of the daily statistics reset enable. Example: DAT:DST:REN 1

The <bool> parameter turns the daily statistics reset on (1 or ON) or off (0 or OFF). This command may be password protected (see Password Commands).

6-26

Page 96
Image 96
Fluke 5020A user manual Data Commands