Fluke 5020A user manual Digital Communications Interface, Interface Commands

Models: 5020A

1 142
Download 142 pages 21.97 Kb
Page 75
Image 75

 

 

Digital Communications Interface

6

 

 

Interface Commands

 

 

 

 

 

*OPT?

Returns configuration options

 

 

 

 

 

 

*RST

Sets the instrument operating parameters to defined

 

 

 

conditions

 

 

 

 

 

 

*SRE?

Returns the Service Request Enable register

 

 

 

 

 

 

*SRE <num>MINMAXDEF

Sets the Service Request Enable register

 

 

 

 

 

 

*STB?

Returns the Status Byte register

 

 

 

 

 

 

*TST? [<bool>]

Returns the self-test results (0: startup test; 1: new

 

 

 

test)

 

 

 

 

 

 

ALARm:BATTery?

Returns the battery alarm event

 

 

 

 

 

 

ALARm:BATTery:ENABle?

Returns the state of the battery alarm enable

 

 

 

 

 

 

ALARm:BATTery:ENABle <bool>

Enables or disables the battery alarm

 

 

 

 

 

 

ALARm:BEEP?

Returns the state of the alarm beep enable

 

 

 

 

 

 

ALARm:BEEP <bool>

Enables or disables the alarm beep

 

 

 

 

 

 

ALARm:CLEar

Clears the alarm events

 

 

 

 

 

 

ALARm:DATE:FIRST?

Returns the date of the first alarm event

 

 

 

 

 

 

ALARm:DATE:LAST?

Returns the date of the latest alarm event

 

 

 

 

 

 

ALARm:DISPlay?

Returns the state of the alarm display enable

 

 

 

 

 

 

ALARm:DISPlay <bool>

Enables or disables the alarm display

 

 

 

 

 

 

ALARm:POWer?

Returns the power loss alarm event

 

 

 

 

 

 

ALARm:POWer:ENABle?

Returns the state of the power loss alarm enable

 

 

 

 

 

 

ALARm:POWer:ENABle <bool>

Enables or disables the power loss alarm

 

 

 

 

 

 

ALARm:RHUMidity<chn>:LOW?

Returns the state of the low humidity alarm for the

 

 

 

specified channel

 

 

 

 

 

 

ALARm:RHUMidity<chn>:LOW:ENABle?

Returns the state of the low humidity alarm enable for

 

 

 

the specified channel

 

 

 

 

 

 

ALARm:RHUMidity<chn>:LOW:ENABle <bool>

Enables or disables the low humidity alarm enable for

 

 

 

the specified channel

 

 

 

 

 

 

ALARm:RHUMidity<chn>:RATE?

Returns the state of the humidity rate alarm for the

 

 

 

specified channel

 

 

 

 

 

 

ALARm:RHUMidity<chn>:RATE:ENABle?

Returns the state of the humidity rate alarm enable for

 

 

 

the specified channel

 

 

 

 

 

 

ALARm:RHUMidity<chn>:RATE:ENABle <bool>

Enables or disables the humidity rate alarm for the

 

 

 

specified channel

 

 

 

 

 

 

ALARm:RHUMidity<chn>:RATE:LIMit?

Returns the humidity rate alarm limit for the specified

 

 

 

channel

 

 

 

 

 

 

ALARm:RHUMidity<chn>:RATE:LIMit

Sets the humidity rate alarm limit for the specified

 

 

<num>MINMAXDEF

channel

 

 

 

 

 

 

ALARm:RHUMidity<chn>:SENSor?

Returns the state of the humidity sensor alarm for the

 

 

 

specified channel

 

 

 

 

 