Calibration Procedure 7
Temperature and Humidity Adjustments
that the measurement errors first be obtained according to the procedure above in section Temperature and Humidity Measurement Errors. After the Temperature and Humidity Adjustments procedure is completed, perform the Temperature and Humidity Measurement Errors procedure again to verify the calibration.
1.Calculate the adjustment ∆ TSL to the temperature slope parameter TSL, and the new temperature slope parameter:
∆ TSL = −10°C ⋅ [Terror(T 3) − Terror(T1)]
(T 3 − T1)
TSL′ = TSL + ∆ TSL
2. Calculate the
Terror(Tn)′ = Terror(Tn) + ∆ TSL ⋅ (Tn − 25°C)
10°C
3. Calculate the median
Terror(T 3)′ + Terror(T 2)′
2
4.Calculate the adjustment ∆ HSL to the humidity slope parameter HSL, and the new humidity slope parameter:
HSL = −25% ⋅ | [RHerror(RH 3) − RHerror(RH 2)] | |
(RH 3 − RH1) | ||
|
HSL′ = HSL + ∆ HSL
5. Calculate the
RHerror(RHn)′ = H Re rror(RHn) + ∆ HSL ⋅ (RHn − 45%) 25%
6.Calculate the median
| | ′ | ′ | |
| RHerror(RH 3) | + THerror(RH 2) | | |
RHmedian′ = | |
|
| |
|
| 2 |
| |
|
|
|
| |
∆HOS = −RHmedian′
HOS ′ = HOS + ∆ HOS
7.Enter the new values for TOS (TOS′), TSL (TSL′), HOS (HOS′), and HSL (HSL′) into the