Page Code:

3Dh(Lamp Timer), 3Eh(Job Separation

 

Sheet)

The Page Code fields identifies the format and parameters defined for that mode page.

This scanner supports 3Dh(Lamp Timer), 3Eh(Job Separation Sheet).

The page length field specifies the length in bytes of the mode parameters that follow. If the initiator does not set this value to the value that is returned for the page by the MODE SENSE command, the target shall terminate the command with CHECK CONDITION status. The sense key shall be set to ILLEGAL REQUEST with the additional sense code set to INVALID FIELD IN PARAMETER LIST. The target is permitted to implement a mode page that is less than the full page length defined by this standard provided no field is truncated and the page length field correctly specifies the actual length implemented.

(a) Lamp Timer page

The Lamp Timer page provides to adjust the Turn-Off time of fluorescent lamp.

Table 3.72 Lamp Timer page

 

 

 

Bit

 

7

 

 

 

6

5

 

4

 

3

 

2

1

0

 

 

 

Byte

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

0

 

 

 

 

 

 

Reserved

 

 

 

Page Code (3Dh)

 

 

 

 

 

 

 

 

1

 

 

 

 

 

 

 

 

 

 

 

 

Page Length (06h)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2

 

 

 

 

 

 

 

 

 

 

 

 

 

Lamp Timer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3

 

 

 

 

 

 

 

 

 

 

 

 

 

Reserved

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

7

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Lamp Timer:

 

 

0 to 255 (second)

 

 

 

 

 

 

 

 

 

This field specifies the fluorescent lamp turn off time in second after scanning. 0 specifies default time(60 seconds). 1 through 255 seconds can be specified.

3-90

M3099GX/GH OEM Manual