Termination
All signals not defined as RESERVED, GROUND, or TERMPWR shall be terminated at both ends of the cable. The Implementor may choose one of the following two methods to terminate each end (see ANSI
INTERFACE
Single-ended alternative (a) [Passive terminator]
TERMINATOR POWER
220 Ω | |
|
330Ω
GROUND
Figure 3.2 Alternative 1 termination for Single Ended Devices
Single-ended alternative (b) [Active terminator]
|
|
|
|
| R3 |
Vin | Low dropout voltage regulater | Vout |
|
| |
Imax ≥ 600mA |
|
| |||
+ | Vout = 2.85V |
| + |
| |
(See Note 1) | R1 | C3 | |||
|
| Vadj |
| C2 | |
|
|
|
|
| |
Component | Description | R2 |
|
| |
R1 | 121 | Ω, 1%, 0.25W |
|
| |
|
|
| |||
R2 | 154 | Ω, 1%, 0.25W |
|
|
|
110 | Ω, 1% |
|
|
| |
C1 | 10 μF Alum.15V or 4.7 μF Tant. 15V |
| R20 | ||
C2 | 150 | μF Alum.10V or 22 μF Tant. 10V (ESR at 120 Hz < 4) |
| ||
|
| ||||
C3 | 0.1 μF Ceramic 25V |
|
|
|
NOTES
1 The voltage regulator shown is an adjustable type with Vref=1.25V. R1 and R2 were selected to provide approximately 10mA Iquiescent. The voltage regulator Vdropout shall be 1.25V or less at Imax.
2 Alternative values that provide lower performance at somewhat lower cost use Vout 1%; R3−R20=100 Ω,2%
Figure 3.3 Alternative 2 termination for Single Ended Devices
The first termination method above is the same as in
M3099GX/GH OEM Manual |