Part # JIBSM08 Rev.6 (08/04/09) Page 29
PROGRAMMING continued
8. Press to return to “PROGRAMMING MODE SYSTEM
SETUP”
9. Press again to exit.
To Change the Alarm Volume:
1. Press and hold for three, (3) seconds. “ENTER CODE” is
displayed.
2. Using the Product buttons, 0-9 enter the code, (1251).
“ENTER CODE **** ” is displayed.
3. Press to enter the Programming Mode.
“PROGRAMMING MODE MENU ITEMS” is displayed.
4. Press two, (2) times. “PROGRAMMING MODE SYSTEM
SETUP” is displayed.
5. Press three, (3), times. “ALARM VOLUME LOW (or
HIGH)” is displayed.
6. Use or to change the Alarm Volume to either Low
or High.
7. Press to save the changes.
8. Press to return to “PROGRAMMING MODE SYSTEM
SETUP”
9. Press again to exit.
To Change the Key Chirp:
To turn key chirp on or o in programing mode only.
1. Press and hold for three, (3) seconds. “ENTER CODE” is
displayed.
2. Using the Product buttons, 0-9 enter the code, (1251).
“ENTER CODE **** ” is displayed.
3. Press to enter the Programming Mode.
“PROGRAMMING MODE MENU ITEMS” is displayed.
4. Press two, (2) times. “PROGRAMMING MODE SYSTEM
SETUP” is displayed.
5. Press four, (4), times. “KEY CHIRP ON (or OFF” is
displayed.
6. Use or to change the key chirp to either ON or OFF.
7. Press to save the changes.
8. Press to return to “PROGRAMMING MODE SYSTEM
SETUP”
9. Press again to exit.
To Change Probe Calibration - Upper:
1. Press and hold for three, (3) seconds. “ENTER CODE” is
displayed.
2. Using the Product buttons, 0-9 enter the code, (1251).
“ENTER CODE **** ” is displayed.
3. Press to enter the Programming Mode.
“PROGRAMMING MODE MENU ITEMS” is displayed.
4. Press two, (2) times. “PROGRAMMING MODE SYSTEM
SETUP” is displayed.
5. Press five, (5), times. “PROBE CAL – UPPER XXXF” is
displayed.
6. Use or to change the Probe Calibration to the
correct temperature measured with a pyrometer.* SEE
CALIBRATION REQUIREMENTS and CALIBRATION section.
7. Press to save the changes.
8. Press to return to “PROGRAMMING MODE SYSTEM
SETUP”
9. Press again to exit.
To Change Probe Calibration - Grill:
1. Press and hold for three, (3) seconds. “ENTER CODE” is
displayed.
2. Using the Product buttons, 0-9 enter the code, (1251).
“ENTER CODE **** ” is displayed.
3. Press to enter the Programming Mode.
“PROGRAMMING MODE MENU ITEMS” is displayed.
4. Press two, (2) times. “PROGRAMMING MODE SYSTEM
SETUP” is displayed.
5. Press six, (6), times. “PROBE CAL – GRILL XXXF” is
displayed.
6. Use or to change the Probe Calibration to the
correct temperature measured with a pyrometer.* SEE
CALIBRATION REQUIREMENTS and CALIBRATION Section.
7. Press to save the changes.
8. Press to return to “PROGRAMMING MODE SYSTEM
SETUP”
9. Press again to exit.
*CALIBRATIONREQUIREMENTS
The control software will beep to the user when it is
appropriate to calibrate. The control will allow calibration
provided the following conditions are met: