ADJUSTMENTS

MODEL 8559A

 

ADJUSTMENTS

5-18. CALIBRATED SWEEP TIME ADJUSTMENT

REFERENCE:

A9 Schematic

DESCRIPTION:

A counter is used to adjust the time interval of the 1 millisecond per division and 5 milliseconds per division sweep times. Calibrated sweep times from 0.1 milliseconds through 50 milliseconds are then checked using the counter time-interval (T.I.) function.

D l SPLAY

COUNTER

0 MHz

SPECTRUM

ANALYZER

FIGURE 5-2.CALIBRATEDSWEEP TIME ADJUSTMENT TEST SETUP

EQUIPMENT:

 

50 MHz Universal Counter

.......................................... H P 5300B/5302A

Extender Cable Assembly

.............................................. H P 5060-0303

BNC Tee ............................................................

H P 1250-0781

PROCEDURE:

 

 

NOTE

Since the calibrated sweep time adjustments are dependent on the + 14.5V and - 10V power supplies, the Power Supply Checks and Adjustments (par- agraph 5-17) should be performedbefore starting this procedure.