MODEL 8559A

ADJUSTMENTS

ADJUSTMENTS

5-20. 1-dB OFFSET ADJUSTMENT(Cont'd)

2.Connect equipment as shown in Figure 5-4.

3.Adjust vertical position to align trace on bottom graticule.

4.Set tuning to 35 MHz. Set TUNING control to center the trace on the display. Set REF LEVEL FINE for a full-screen trace (signal at top graticule line).

5.Set Amplitude Scale to 10 dB/DIY Adjust VERT GAIN if necessary for full screen trace.

6 . Repeat steps 3 and 4 until the trace is full screen in both LIN and 10 dB/DIY

NOTE

1 dBlDlV will read approximately 0.5 dB (0.5 division) low when using extender cable assembly. Adjusting A15R11 dB OFFSET for a trace 0.5 division down from top graticule line should place signal at top graticule line when HP 8559A is properly installed in display mainframe.

7. Set Amplitude Scale to 1 dB/DIV. Adjust A15R1 1 dB OFFSET for a trace 0.5 division down from top graticule line.